Hanging-tube instruments: Start the flushing cycle by pressing
Instrument spray (you will hear a signal tone) and holding it for 6
seconds. Release the button when you have heard a second and a third
signal tone.
The system will first identify the instruments and then each instrument
hose is flushed with air and water in its turn. The time of flushing is the
same for all instruments. The total flushing time is displayed on the
control panel. At the same time the cup fill tube and the bowl are flushed,
too.
NOTE
Adjust the water flow of the cup fill tube so that the water does not splash
into the bowl.
5. Once the flushing cycle is finished, H 36 is displayed. Remove the
instruments from the flushing holder and return them to the instrument
console. Place the flushing holder cover back to its position.
If water flow was not detected for all water consuming instruments,
H 36.1 is displayed. Run instruments for a while with spray water to
ensure that the flushing of instruments is sufficient.
Press Instrument spray to view the instrument flushing results on the
display. The bars on the display represent the instruments in the order
they are placed in the instrument console. A full bar means that the
instrument was flushed successfully and a half bar that the flushing for
that instrument failed.
What to do next
NOTE
After the flushing cycle, the unit should immediately be switched off. This
procedure ensures that cool water remains in the pipelines of the unit, thus
minimising the growth of biofilm.
NOTE
The dental unit can be configured to display help message H 99 if the unit
has been switched off without performing long flushing after water has been
used. When H 99 has been enabled, it can be displayed as a reminder only,
or it can be set to disappear only after you have performed long flushing. To
take help message H 99 into use, contact your Planmeca dealer.
23.5 Suction cleaning
About this task
NOTE
The Suction Tube Cleaning System (STCS) is an optional feature. If your
dental unit does not have this feature, clean the suction tubes with OroCup.
For more information, see section "OroCup" on page 125.
Steps
1. Remove the suction handpieces from the suction tubes and clean them
according to the manufacturer’s instructions.
2. Open the suction cleaning cover.
23 Flushing and cleaning programs
User's manual
Planmeca Compact i 115
Содержание Compact i Classic v2
Страница 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Страница 8: ...Table of contents Planmeca Compact i User s manual...
Страница 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Страница 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Страница 185: ...30 Technical information User s manual Planmeca Compact i 177...
Страница 189: ......