Remove the supplementary holder by pulling it out from the Flexy-holder. To
replace it, push it firmly into its place.
Remove the holder for the intraoral scanner by pressing the trigger (see
arrow in picture) and pulling the holder out from the Flexy-holder. To replace
the holder for the intraoral scanner, push it firmly into its place.
11.3 Removing and replacing suction tubes
Steps
1. To minimise contamination risk, perform suction cleaning.
For instructions, see section "Suction cleaning" on page 115.
2. Remove the side cover plate by pulling it out from the cuspidor.
3. Detach the securing ring (1) from the suction tube.
NOTE
You can grease the securing ring with non-toxic vaseline to make it
easier to remove.
11 Suction system
42 Planmeca Compact i
User's manual
Содержание Compact i Classic v2
Страница 1: ...PlanmecaCompact iClassicv2 user s manual EN 10032684...
Страница 8: ...Table of contents Planmeca Compact i User s manual...
Страница 23: ...8 Planmeca Compact i dental unit User s manual Planmeca Compact i 15...
Страница 114: ...4 Press Program 22 Programming 106 Planmeca Compact i User s manual...
Страница 185: ...30 Technical information User s manual Planmeca Compact i 177...
Страница 189: ......