INDEX MATCHING GEL CARTRIDGE REPLACEMENT

The 1120 Index Matching Cartridges contain enough matching compound 

which, under normal use, should last for over 500 fiber insertions.  Inconsistent 

couplings or high coupling losses or reflections may indicate the need to refill 

(112X-RMC) or replace (112X-IMC) the cartridge.  

In order to replace the index matching gel cartridge, hold the coupling module 

and unscrew its end cap (6).  Pull the empty cartridge (5) from the sleeve (4) 

and discard.  Insert a new cartridge by first removing and discarding the sealing 

tabs from the ends of the IMC replacement cartridge, then simply push the large 

end of cartridge into coupling module sleeve and screw on the end cap until it is 

finger tight.

INDEX MATCHING GEL CARTRIDGE (112X-RMC) REFILLING

If you have purchased the refillable version of the index matching gel cartridge, 

the 112X-RMC, you can replenish the gel in your index matching gel cartridge 

with the gel-filled syringe (112X-SYR) that is purchased separately or included 

as part of the 112X-RFK Refill Kit.   

To refill the RMC, remove it from the 1120 Coupling Module as described in the 

previous section.  Next, unscrew the RMC cap and then, after removing the 

syringe dust cap, attach the RMC to the syringe’s tip adapter.  Inject index 

matching gel into the RMC until gel starts to emerge from the opposite end of 

the cartridge indicating that the cartridge is full.  Remove the RMC from the 

syringe and replace its cap.  Clean any excess gel from the cartridge with a lint 

free wipe. The RMC is now ready to be loaded into the 1120 coupling module.  

Store the syringe in an appropriate cool, dust-free location to prevent leakage or 

contamination.

5

Refillable Index Matching Cartridge with 112X-SYR Syringe

Syringe with RMC Attached

Содержание 1120

Страница 1: ...SA tel 1 503 644 1960 fax 1 503 526 4700 Technical Response Center tel 1 503 526 4678 Technical Response Center fax 1 503 526 4633 Technical Response Center e mail support pkinetics com www pkinetics com 90 0588 00 Rev H July 2019 1120 BARE FIBER ALIGNER USER S GUIDE ...

Страница 2: ...econditioned parts that fully meet applicable specifications The Warranty for these replacement parts is thirty 30 days or the remainder of the Warranty period whichever is longer b Products Covered The Warranty covers Products as delivered by Photon Kinetics to the customer in unmodified condition The customer understands that modification of any Product without Photon Kinetics prior written cons...

Страница 3: ... to an optical source such as an OTDR as the total light power emitted from the light source may also be present at the output of the bare fiber aligner or at the output of an optical fiber inserted into the bare fiber aligner Warning Handling Optical Fibers Exercise care when handling the unprotected ends of optical fibers use appropriate eye protection and dispose of discarded fiber ends in a sa...

Страница 4: ...ssemble a complete aligner from the components shown below Several different configurations are possible by combining either a Field or Production Aligner 1120F or 1120P with an 1120B Buffer Fiber or 1120J Jumper Fiber These compo nents may also be purchased individually Illustrations of each of the four pro duction configurations are shown below 1 1120M Production Aligner Module Complete Module 1...

Страница 5: ...0P sim ply pull the gray cylindrical end cap from the housing some initial resistance will be encountered as the cap is restrained by an internal spring loaded ball plunger BEFORE YOU BEGIN Load the the 112X IMC or 112X RMC Index Matching Cartridge into the 1120 Bare Fiber Aligner by following the instructions below Unscrew and remove the coupling module end cap to expose the internal metal ferrul...

Страница 6: ...established completely remove the fiber and re prepare the fiber end SPECIFICATIONS Coupling Loss 0 5 dB typical Coupling Reflectance 45 dB typical Operating Temperature 10 to 30 C Dimensions Buffer Fiber Module 1120B Field Aligner 1120F Production Module 1120M Production Aligner 1120P 12 5 x 12 5 x 2 5 cm 4 9 x 4 9 x 1 0 in 3 1 x 3 7 x 2 4 cm 1 2 x 1 5 x 0 9 in 11 9 x 16 3 x 4 9 cm 4 7 x 6 4 x 1 ...

Страница 7: ...r the cartridge can be cleaned with 100 isopro panol Complete immersion in an ultrasonic bath is optimal If a fiber happens to break off inside the ferrule assembly use the 112X CLK Ferrule Cleaning Kit to clear the fragment Simply insert the cleaning ferrule included in the kit into the ferrule assembly 2 3 4 5 as shown below left and then feed a cleaning wire into the ferrule barrel and through ...

Страница 8: ...ve purchased the refillable version of the index matching gel cartridge the 112X RMC you can replenish the gel in your index matching gel cartridge with the gel filled syringe 112X SYR that is purchased separately or included as part of the 112X RFK Refill Kit To refill the RMC remove it from the 1120 Coupling Module as described in the previous section Next unscrew the RMC cap and then after remo...

Страница 9: ...arton Inadequate packaging may lead to serious damage and may invalidate your remaining warranty Before returning the 1120 you must obtain a Return Materials Authorization RMA number from Photon Kinetics To obtain an RMA number please have your model number and serial number available and call 1 503 526 4678 or send an email to support pkinetics com All aligners should be returned to the following...

Страница 10: ...7 INDEX MATCHING GEL MATERIAL SAFETY DATA SHEET ...

Страница 11: ...8 ...

Страница 12: ... Matching Cartridge pack of 3 112X IMC 12 Disposable Index Matching Cartridge pack of 12 112X CLK Ferrule Cleaning Kit wires and cleaning ferrule 112X CLW Ferrule Cleaning Wires 112X RFB Re termination of the FC PC end 112X RSS Replacement Sleeve Set 112X RFA Replacement Ferrule Assembly Photon Kinetics Inc 9305 S W Gemini Drive Beaverton Oregon 97008 USA tel 1 503 644 1960 fax 1 503 526 4700 Tech...

Отзывы:

Похожие инструкции для 1120