re-insert it back into the ceramic sleeve. Re-secure the ferrule by rotating the 

retaining ring (2) clockwise, being careful not to overtighten the ring. Note that 

over time the connector will wear, but they can be refurbished. Replacement 

jumper cables and jumper/buffer fiber assemblies are also available. 

For more thorough cleaning, remove the coupling module assembly and gel car-

tridge (2-7) from the 1120F base (8) or the 1120M/P end cap (9) by unscrewing 

the adapter (4) from the base/end cap. Next, loosen the retaining ring (2) and 

pull apart the ferrule assembly (5,6) and gel cartridge (7) should slide out of the 

end cap/base (some gentle pressure on the end of the gel cartridge may be 

required). All parts, except for the cartridge, can be cleaned with 100% isopro-

panol. Complete immersion in an ultrasonic bath is optimal.

If a fiber happens to break off inside the ferrule assembly use the 112X-CLK 

Ferrule Cleaning Kit to clear the fragment. Simply insert the cleaning ferrule 

included in the kit into the ferrule assembly (2,3,4,5) as shown below left, and 

then feed a cleaning wire into the ferrule barrel and through both the cleaning 

and aligner ferrules to push out any fiber fragments.

If a cleaning ferrule is not available, it is possible to use just a cleaning wire to 

clear fiber fragments. First, disassemble the coupling module to access the 

aligner ferrule (5). Then grip a cleaning wire very close to its end (3-4 mm) and 

feed the wire into the ferrule as shown below. Note that some initial resistance 

may be encountered due to the recess surrounding the ferrule bore (right).  

4

Coupling Module Assembly

8

9

Содержание 1120

Страница 1: ...SA tel 1 503 644 1960 fax 1 503 526 4700 Technical Response Center tel 1 503 526 4678 Technical Response Center fax 1 503 526 4633 Technical Response Center e mail support pkinetics com www pkinetics com 90 0588 00 Rev H July 2019 1120 BARE FIBER ALIGNER USER S GUIDE ...

Страница 2: ...econditioned parts that fully meet applicable specifications The Warranty for these replacement parts is thirty 30 days or the remainder of the Warranty period whichever is longer b Products Covered The Warranty covers Products as delivered by Photon Kinetics to the customer in unmodified condition The customer understands that modification of any Product without Photon Kinetics prior written cons...

Страница 3: ... to an optical source such as an OTDR as the total light power emitted from the light source may also be present at the output of the bare fiber aligner or at the output of an optical fiber inserted into the bare fiber aligner Warning Handling Optical Fibers Exercise care when handling the unprotected ends of optical fibers use appropriate eye protection and dispose of discarded fiber ends in a sa...

Страница 4: ...ssemble a complete aligner from the components shown below Several different configurations are possible by combining either a Field or Production Aligner 1120F or 1120P with an 1120B Buffer Fiber or 1120J Jumper Fiber These compo nents may also be purchased individually Illustrations of each of the four pro duction configurations are shown below 1 1120M Production Aligner Module Complete Module 1...

Страница 5: ...0P sim ply pull the gray cylindrical end cap from the housing some initial resistance will be encountered as the cap is restrained by an internal spring loaded ball plunger BEFORE YOU BEGIN Load the the 112X IMC or 112X RMC Index Matching Cartridge into the 1120 Bare Fiber Aligner by following the instructions below Unscrew and remove the coupling module end cap to expose the internal metal ferrul...

Страница 6: ...established completely remove the fiber and re prepare the fiber end SPECIFICATIONS Coupling Loss 0 5 dB typical Coupling Reflectance 45 dB typical Operating Temperature 10 to 30 C Dimensions Buffer Fiber Module 1120B Field Aligner 1120F Production Module 1120M Production Aligner 1120P 12 5 x 12 5 x 2 5 cm 4 9 x 4 9 x 1 0 in 3 1 x 3 7 x 2 4 cm 1 2 x 1 5 x 0 9 in 11 9 x 16 3 x 4 9 cm 4 7 x 6 4 x 1 ...

Страница 7: ...r the cartridge can be cleaned with 100 isopro panol Complete immersion in an ultrasonic bath is optimal If a fiber happens to break off inside the ferrule assembly use the 112X CLK Ferrule Cleaning Kit to clear the fragment Simply insert the cleaning ferrule included in the kit into the ferrule assembly 2 3 4 5 as shown below left and then feed a cleaning wire into the ferrule barrel and through ...

Страница 8: ...ve purchased the refillable version of the index matching gel cartridge the 112X RMC you can replenish the gel in your index matching gel cartridge with the gel filled syringe 112X SYR that is purchased separately or included as part of the 112X RFK Refill Kit To refill the RMC remove it from the 1120 Coupling Module as described in the previous section Next unscrew the RMC cap and then after remo...

Страница 9: ...arton Inadequate packaging may lead to serious damage and may invalidate your remaining warranty Before returning the 1120 you must obtain a Return Materials Authorization RMA number from Photon Kinetics To obtain an RMA number please have your model number and serial number available and call 1 503 526 4678 or send an email to support pkinetics com All aligners should be returned to the following...

Страница 10: ...7 INDEX MATCHING GEL MATERIAL SAFETY DATA SHEET ...

Страница 11: ...8 ...

Страница 12: ... Matching Cartridge pack of 3 112X IMC 12 Disposable Index Matching Cartridge pack of 12 112X CLK Ferrule Cleaning Kit wires and cleaning ferrule 112X CLW Ferrule Cleaning Wires 112X RFB Re termination of the FC PC end 112X RSS Replacement Sleeve Set 112X RFA Replacement Ferrule Assembly Photon Kinetics Inc 9305 S W Gemini Drive Beaverton Oregon 97008 USA tel 1 503 644 1960 fax 1 503 526 4700 Tech...

Отзывы:

Похожие инструкции для 1120