OPERATION

  

The 1120 Bare Fiber Aligner is used to quickly connect bare optical fiber to a 

test equipment optical connector.  Before using the 1120, ensure that a 112X-

IMC or 112X-RMC Index Matching Cartridge is loaded into the 1120’s coupling 

module (see previous section) and that the opposite end of the 1120 is connect-

ed to the OTDR or other test equipment bulkhead.

   

To couple a fiber to the test instrument with the 1120 Bare Fiber Aligner:

1. Strip at least 35 mm of coating from the fiber to expose bare glass and clean 

the fiber with isopropyl alcohol and a lint-free wipe.

2. Either break or cleave the fiber at a point approximately 25 mm from the 

coating shoulder.  

3. Slowly insert the fiber into the end of the 1120 Bare Fiber Aligner coupling 

module until it contacts the module’s internal connector ferrule.  If an optical 

connection is not accomplished on the first insertion, pull the fiber back about 

1/8” and push it forward, again making contact with the connector.  If a connec-

tion still cannot be established, completely remove the fiber and re-prepare the 

fiber end.

 

SPECIFICATIONS

Coupling Loss

< 0.5 dB typical

Coupling Reflectance

< -45 dB typical 

Operating Temperature

10 to 30

°C

Dimensions

Buffer Fiber Module (1120B)

Field Aligner (1120F)

Production Module (1120M)

Production Aligner (1120P)

12.5 x 12.5 x 2.5 cm (4.9 x 4.9 x 1.0 in.)

3.1 x 3.7 x 2.4 cm (1.2 x 1.5 x 0.9 in.)

11.9 x 16.3 x 4.9 cm (4.7 x 6.4 x 1.9 in.)

9.5 x 6.2 x 3.6 cm (3.7 x 2.4 x 1.4 in.)

Index Matching Gel Cartridge Lifetime

112X-IMC

112X-RMC

> 500 fiber couplings (typical)

> 500 fiber couplings per refill (typical)

MAINTENANCE

CLEANING

Occasionally you may have to clean the jumper connector (with red strain relief) 

that is attached to the 1120 coupling module, and/or the ferrule assembly/metal-

lic sleeve inside the 1120 coupling module.  For the 1120M or P aligners, first 

pull the end cap (9) and attached coupling module assembly from the aligner 

housing (do not pull the end cap more than 20 cm out of the 1120M housing).  

To remove the connector (1) from the assembly, loosen the knurled retaining 

ring (2) by rotating it counterclockwise and then pull the connector ferrule out of 

the ceramic sleeve (3) inside the bulkhead housing. Clean the ferrule and then 

3

Содержание 1120

Страница 1: ...SA tel 1 503 644 1960 fax 1 503 526 4700 Technical Response Center tel 1 503 526 4678 Technical Response Center fax 1 503 526 4633 Technical Response Center e mail support pkinetics com www pkinetics com 90 0588 00 Rev H July 2019 1120 BARE FIBER ALIGNER USER S GUIDE ...

Страница 2: ...econditioned parts that fully meet applicable specifications The Warranty for these replacement parts is thirty 30 days or the remainder of the Warranty period whichever is longer b Products Covered The Warranty covers Products as delivered by Photon Kinetics to the customer in unmodified condition The customer understands that modification of any Product without Photon Kinetics prior written cons...

Страница 3: ... to an optical source such as an OTDR as the total light power emitted from the light source may also be present at the output of the bare fiber aligner or at the output of an optical fiber inserted into the bare fiber aligner Warning Handling Optical Fibers Exercise care when handling the unprotected ends of optical fibers use appropriate eye protection and dispose of discarded fiber ends in a sa...

Страница 4: ...ssemble a complete aligner from the components shown below Several different configurations are possible by combining either a Field or Production Aligner 1120F or 1120P with an 1120B Buffer Fiber or 1120J Jumper Fiber These compo nents may also be purchased individually Illustrations of each of the four pro duction configurations are shown below 1 1120M Production Aligner Module Complete Module 1...

Страница 5: ...0P sim ply pull the gray cylindrical end cap from the housing some initial resistance will be encountered as the cap is restrained by an internal spring loaded ball plunger BEFORE YOU BEGIN Load the the 112X IMC or 112X RMC Index Matching Cartridge into the 1120 Bare Fiber Aligner by following the instructions below Unscrew and remove the coupling module end cap to expose the internal metal ferrul...

Страница 6: ...established completely remove the fiber and re prepare the fiber end SPECIFICATIONS Coupling Loss 0 5 dB typical Coupling Reflectance 45 dB typical Operating Temperature 10 to 30 C Dimensions Buffer Fiber Module 1120B Field Aligner 1120F Production Module 1120M Production Aligner 1120P 12 5 x 12 5 x 2 5 cm 4 9 x 4 9 x 1 0 in 3 1 x 3 7 x 2 4 cm 1 2 x 1 5 x 0 9 in 11 9 x 16 3 x 4 9 cm 4 7 x 6 4 x 1 ...

Страница 7: ...r the cartridge can be cleaned with 100 isopro panol Complete immersion in an ultrasonic bath is optimal If a fiber happens to break off inside the ferrule assembly use the 112X CLK Ferrule Cleaning Kit to clear the fragment Simply insert the cleaning ferrule included in the kit into the ferrule assembly 2 3 4 5 as shown below left and then feed a cleaning wire into the ferrule barrel and through ...

Страница 8: ...ve purchased the refillable version of the index matching gel cartridge the 112X RMC you can replenish the gel in your index matching gel cartridge with the gel filled syringe 112X SYR that is purchased separately or included as part of the 112X RFK Refill Kit To refill the RMC remove it from the 1120 Coupling Module as described in the previous section Next unscrew the RMC cap and then after remo...

Страница 9: ...arton Inadequate packaging may lead to serious damage and may invalidate your remaining warranty Before returning the 1120 you must obtain a Return Materials Authorization RMA number from Photon Kinetics To obtain an RMA number please have your model number and serial number available and call 1 503 526 4678 or send an email to support pkinetics com All aligners should be returned to the following...

Страница 10: ...7 INDEX MATCHING GEL MATERIAL SAFETY DATA SHEET ...

Страница 11: ...8 ...

Страница 12: ... Matching Cartridge pack of 3 112X IMC 12 Disposable Index Matching Cartridge pack of 12 112X CLK Ferrule Cleaning Kit wires and cleaning ferrule 112X CLW Ferrule Cleaning Wires 112X RFB Re termination of the FC PC end 112X RSS Replacement Sleeve Set 112X RFA Replacement Ferrule Assembly Photon Kinetics Inc 9305 S W Gemini Drive Beaverton Oregon 97008 USA tel 1 503 644 1960 fax 1 503 526 4700 Tech...

Отзывы:

Похожие инструкции для 1120