36 -
STR561 - User manual
5
Full scale
Full scale value of strain-gauge sensor, used during calibration type
“
Calibration on full scale % value
”.
-999999
[digit
1
],
Default
: 1000.
6
Lower limit
Lower limit of sensor, used during input calibration type
Sampling value
or
Calibration on full scale % value
.
-999999
[digit
1
],
Default
: 0.
7
Upper limit
Upper limit of sensor, used during input calibration type
Sampling value
or
Calibration on full scale % value
-999999
[digit
1
],
Default
: 1000.
8
Offset calibration
Analogue input offset correction. Value added / subtracted to the process
visualization.
-10000..+10000
[digit
1
],
Default
0.0.
9
Gain calibration
Percentage value that is multiplied for the process value (allows to
calibrated the working point)
-100.0%..+100.0%,
Default
: 0.0
ex: to correct the range from 0..1000kg showing 0..1010kg, set the
parameter to -1.0.
10 Calibration
Selection of calibration type for the analogue input. After finishing or
deleting calibration proceeding, this parameter will automatically return
to default setting (=Disabled).
Disabled (
Default
)
Full scale % value
Sampling value
mV/V Value