Operating Instructions
STORING HI and LO EZ-CHECK Source Outputs
Speed up your calibration by storing Span & Zero output setting
for instant recall with the EZ-CHECK switch.
1) Store your high (SPAN) output temperature by moving the
EZ-CHECK switch to the
HI
position and turning the EZ-Dial
knob until the desired output value is on the display. Press
and hold the EZ-Dial knob until
STORED
appears to store
the value. Release the EZ-Dial knob.
2) Store your low (ZERO) output value by moving the EZ-CHECK
switch to the
LO
position and turning the EZ-Dial knob until
the desired output value is on the display. Press and hold the
EZ-Dial knob until
STORED
appears to store the value.
Release the EZ-Dial knob.
3) Instantly output your SPAN and ZERO output settings by
moving the EZ-CHECK switch between HI and LO. You may
also select any third output setting (such as mid-range) using
the SET position on the EZ-CHECK switch.
CHANGING BATTERIES
Low battery is indicated by a battery symbol on the display.
Approximately one to four hours of typical operation remain before
the PIE 525Plus will automatically turn off. To change the batteries
remove the rubber boot and remove the battery door from the
back of the unit by sliding the door downward. This allows access
to the battery compartment. Replace with four (4) “AA” 1.5V
batteries being careful to check the polarity. Replace the battery
door and replace the boot. All stored configuration options (T/C
Type, EZ-CHECK Memories, etc.) are reset to factory settings
when the batteries are removed.
Note:
Alkaline batteries are supplied and recommended for
typical battery life and performance. Optional rechargeable
batteries (charged externally) are available.