Page 20
Source Resistance
Choose this function to provide a simulated resistance into any
device that measures resistance.
Move the power switch
w
to SOURCE then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select OHMS for the FUNCTION, 400Ω or 4000Ω for the RANGE.
Turn on SENSOR mA to check the excitation current generated
by the receiver.
Disconnect all sensor wires from the devices to be calibrated and
connect the PIE 525Plus to the inputs of the device using 2, 3 or 4 wires.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted for
400Ω/4000Ω ranges in 0.01Ω/0.1Ω increments in LOW and
0.001Ω/0.01Ω increments in HIGH resolution by turning the knob
e
. Press and turn the knob for faster dialing with 1.00Ω/10.0Ω
increments in LOW and 1.000Ω/10.00Ω increments in HIGH
resolution.
2, 3 & 4 Wire Resistance Connections
Instrument with
Resistance Input
Controller
Transmitter
PLC
Ω
HI
01.05mA FIXED
305.25
3 2 1 4
+
-
-
+
OFF
OVERLOAD
Double Click
Menu
HI
SET
LO
SOURCE
MAX
READ
MIN
READ
Push & Hold
to
Store/Step
Push & Turn
for
Fast Dialing
Diagnostic Thermocouple
RTD & Milliamp Calibrator
2, 3, 4 Wire Detection
Loop Leak Detect ∙ Transmitter Supply
www.
GlobalTestSupply
.com
Find Quality Products Online at: