Page 21
Read Resistance
Choose this function to measure resistance.
Move the power switch
w
to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select OHMS for the FUNCTION, 400Ω or 4000Ω for the
RANGE.
Connect the PIE 525Plus to the resistor or sensor using 2, 3 or 4
wires. The PIE 525Plus automatically detects how many wires
are connected using a patented circuit and indicates each wire
that is connected. Any wires that are not connected or broken are
indicated by the 525Plus. This is useful for troubleshooting the
sensor.
Signals above the maximum scale are limited by protection
circuitry with “OVER RANGE” on the display.
The PIE 525Plus measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch
q
to MAX to see the highest reading and to
MIN to see the lowest reading. Press and hold the knob
e
to
clear the MAX and MIN readings.
Ω
250.025
OFF
OVERLOAD
Double Click
Menu
HI
SET
LO
SOURCE
MAX
READ
MIN
READ
Push & Hold
to
Store/Step
Push & Turn
for
Fast Dialing
Diagnostic Thermocouple
RTD & Milliamp Calibrator
2, 3, 4 Wire Detection
Loop Leak Detect ∙ Transmitter Supply
2 1
+
-
2, 3 & 4 Wire Resistance Connections
www.
GlobalTestSupply
.com
Find Quality Products Online at: