Page 12
Double click the
e
DIAL KNOB at any time the unit is on and
then turn the
e
DIAL KNOB to move to the second menu page
so the word
DISPLAY
appears at the top of the menu.
Turn the
e
DIAL KNOB to move through the menu. Press the
e
DIAL KNOB to toggle between LOW and HIGH or OFF and ON.
LOW resolution is 0.001 mV, 0.01Ω in 400Ω Range, 0.1Ω in
4000Ω Range and 0.1° for T/C & RTD. HIGH resolution is 0.0001
mV, 0.001Ω in 400Ω Range, 0.01Ω in 4000Ω Range and 0.01° for
T/C & RTD. Milliamp always has resolution of 0.001 mA/0.01%.
V DISPLAY
> EXIT
(2/4 SOURCE, 2/3 READ)
RESOLUTION LOW
HIGH
Operating Instructions
Double Click Menu - DISPLAY Page
OHMS DISPLAY
> EXIT
(2/4 SOURCE, 2/3 READ)
RESOLUTION LOW
HIGH
SENSOR mA* OFF
ON
RTD DISPLAY
> EXIT
(2/4 SOURCE, 2/3 READ)
RESOLUTION LOW
HIGH
DISPLAY OHMS OFF
ON
SENSOR mA* OFF
ON [Source Ohm & RTD Only]
T/C DISPLAY
> EXIT
(2/4 SOURCE, 2/3 READ)
RESOLUTION LOW
HIGH
DISPLAY mV OFF
ON
DISPLAY CJ
OFF
ON [Cold Junction Temperature]
TC PROBE
Ω
OFF
ON [Probe Resistance - Read Only]
EXIT MENU
- exits this menu immediately and saves any
changes. Menu will automatically exit after 15 seconds of
inactivity. Settings are remembered even with the power off.