Diagnostic Thermocouple
RTD & Milliamp Calibrator
2, 3, 4 Wire Detection
Loop Leak Detect ∙ Transmitter Supply
Milliamp Output Signal
Controller
Transmitter
P/I
DCS
mA
19.997
5 6
+
-
HART
READ
READ mA, READ % (Percent of 4 to 20 mA)
Choose this function to measure from 0.000 to 24.000 milliamps
or -25.00 to 125.00%.
Move the power switch
w
to READ then Double Click the
EZ-DIAL knob to get into the Menu. Turn the knob to scroll
through the settings and press the knob to make your selection.
Select
mA
for the
FUNCTION
and
READ
for the
MODE
.
Choose either
mA
or
%
and whether you need the 250
Ω
HART
resistor active in the loop.
Connect the red input lead (+) of the PIE 525Plus to the more
positive point of the break and the black input to the more
negative point.
Signals below 0 mA or open circuits are indicated by 0.000 mA
(-25.00%) on the display. Signals above 24 mA are current
limited by protection circuitry with “OVERRANGE” flashed on the
display and the red OVERLOAD LED lit.
The PIE 525Plus measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch
q
to MAX to see the highest reading and to
MIN to see the lowest reading. Press and hold the knob
e
to
clear the MAX and MIN readings.