Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

AUTO OFF

 - If 

AUTO OFF

 is 

ON

, the unit will 

turn off after 30 minutes of inactivity to save 
battery life. If 

AUTO OFF

 is 

OFF

 the unit will 

stay on until the POWER SWITCH is moved to 
the off position.

BACKLIGHT

 - If 

BACKLIGHT

 is 

ON

 the 

backlight will light all the time the unit is powered 
up. For maximum battery life turn the backlight 
off when using the calibrator in areas with 
enough ambient light to read the display.

STEPS/RAMP

 - pressing the knob will cycle 

through 

2, 3, 5, 11

 and 

RAMP

. The endpoints 

of the steps or ramp are based on the values 
stored in the 

HI

 and 

LO

 EZ-CHECK outputs. 

2

 steps

 will automatically switch between the 

values stored in the HI & LO EZ-CHECK (0 & 
100%). 

 

3

 steps

 between the HI, Midpoint and LO 

EZ-CHECK (0, 50 & 100%). 

5

 steps

 between the HI and LO EZ-CHECK 

in 25% increments (0, 25, 50, 75 & 100%).

11

 steps

 between the HI and LO EZ-CHECK 

in 10% increments (0, 10, 20...80, 90 &100%).

RAMP

 continuously ramps up and down 

between the HI and LO EZ-CHECK outputs.

STEP/RAMP TIME

 - pressing the knob will 

cycle through 

5, 6, 7, 8, 9, 10, 15, 20, 25, 

30

 and 

60

 seconds. 

           

FEATURES

> EXIT (4/4)   

      

   AUTO OFF  

        ON 

OFF

   BACKLIGHT   

   ON 

OFF

   STEPS/RAMP         2

 

3 5 11 RAMP

   STEP/RAMP TIME  5 

6 7 8 9 10 15 20 25 30 60

Double Click Menu - FEATURES

Double Click Menu - mA DISPLAY

MODE

 - pressing the knob will cycle through 

READ

PWRM

READ%

PWRM%

 and 

OFF

READ

 turns on the mA display and indicates 

current passing through the loop proportional 
to the input of the transmitter which is controlled 
by the output of the 311Plus. Choose 

READ%

 

to display in percent of 4-20 milliamps.

 

PWRM

 is POWER MEASURE which uses the 

internal loop supply of the 311Plus to power up 
the transmitter while indicating the current 
passing through the loop proportional to the 
input of transmitter which is controlled by the 
output of the 311Plus. Choose 

PWRM%

 to 

display in percent of 4-20 milliamps.

If 

PWRM

 or 

PWRM%

 is selected an 

additional menu selection of 

LEAK DETECT

 

will appear. When 

LEAK DETECT

 

is turned 

on the 311Plus will display 

LEAK

 and the 

amount of current that is uncontrolled in the 
loop. This may be due to a faulty transmitter, 
corrosion causing a bridge to ground or 
moisture present at some connection point. 
When 

LEAK DETECT

 tests a loop with 

leakage the loop mA signal will be affected.

HART 250

Ω

 - turn on the 250

Ω

 

resistor if you 

are powering up a HART transmitter. This 
provides the loop load required for HART 
communication.

     mA DISPLAY (For RTD)

>  EXIT (3/4)   

 

 

      

 

MODE    OFF 

READ PWRM READ% PWRM%

 

HART 250

Ω

     ON 

OFF

 

LEAK DETECT OFF 

ON

Configuration

Содержание 311Plus

Страница 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Страница 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Страница 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Страница 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Страница 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Страница 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Страница 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Страница 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Страница 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Страница 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Страница 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Страница 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Страница 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Страница 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Страница 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Страница 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Страница 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Страница 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Страница 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Страница 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Отзывы: