Operating Instructions

Milliamp Receiver Input

Controller

Transmitter

Computer

Logger

I/P

DCS

OFF

OVERLOAD

Double Click 

Menu

HI

SET

LO

SOURCE

MAX
READ

MIN

READ

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Diagnostic RTD &

Milliamp Calibrator

2, 3, 4 Wire Detection

Loop Leak Detect ∙ Transmitter Supply

mA

20.000

5      6

  

    

+

-

HART

SOURCE

HI

mA SOURCE/  % SOURCE (Percent of 4 to 20 mA)

Choose this function to provide an output from 0.000 to 24.000 milliamps. The compliance voltage 
is a nominal 24 VDC to provide the driving power to your milliamp receivers.

Move the power switch 

to SOURCE then Double Click the EZ-DIAL knob to get into the menu. 

Turn the knob to scroll through the settings and press the knob to make your selection. Select 

mA

 

for the 

FUNCTION

 and 

SOURCE

 for the 

MODE

. Choose either 

mA

 or 

%

 and whether you need 

the 250

 HART resistor active in the loop.

Connect the output leads of the PIE 311Plus to the inputs of the device being calibrated, making 
sure to check polarity. Red lead to the plus (+) input and black lead to the minus (-) input.

Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI 
and LO (defaults to 20 & 4 mA). You may also select any third output setting (such as mid-range) 
using the SET position on the EZ-CHECK switch. The output is adjusted in 0.001 mA (0.01%) 
increments by turning the knob 

e

. Press and turn the knob for faster dialing with 0.100 mA (1.00%) 

increments.

OFF

OVERLOAD

Double Click 

Menu

HI

SET

LO

SOURCE

MAX
READ

MIN

READ

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Diagnostic RTD &

Milliamp Calibrator

2, 3, 4 Wire Detection

Loop Leak Detect ∙ Transmitter Supply

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

mA

19.997

5      6

  

    

+

-

HART

READ

READ mA, READ % (Percent of 4 to 20 mA)

Choose this function to measure from 0.000 to 24.000 milliamps or -25.00 to 125.00%.

Move the power switch 

to READ then Double Click the EZ-DIAL knob to get into the Menu. Turn 

the knob to scroll through the settings and press the knob to make your selection. Select 

mA

 for the 

FUNCTION

 and 

SOURCE

 for the 

MODE

. Choose either 

mA

 or 

%

 and whether you need the 

250

 HART resistor active in the loop.

Connect the red input lead (+) of the PIE 311Plus to the more positive point of the break and the 
black input to the more negative point.

Signals below 0 mA or open circuits are indicated by 0.000 mA (-25.00%) on the display. Signals 
above 24 mA are current limited by protection circuitry with “OVERRANGE” flashed on the display 
and the red OVERLOAD LED lit.

The PIE 311Plus measures the input signal and constantly updates the display with the current 
reading. Move the EZ-CHECK switch 

to MAX to see the highest reading and to MIN to see the 

lowest reading. Press and hold the knob 

e

 to clear the MAX and MIN readings.

All Milliamp Connections

Содержание 311Plus

Страница 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Страница 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Страница 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Страница 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Страница 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Страница 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Страница 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Страница 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Страница 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Страница 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Страница 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Страница 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Страница 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Страница 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Страница 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Страница 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Страница 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Страница 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Страница 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Страница 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Отзывы: