Operating Instructions

OHMS SOURCE

Choose this function to provide an output from 0.00 to 4010.0 ohms. 

Move the power switch 

to SOURCE then Double Click the EZ-DIAL knob to get into the Menu. 

Turn the knob to scroll through the settings and press the knob to make your selection. Select 

OHMS

 for the 

FUNCTION

 and 

400Ω 

or

 4000Ω

 for the 

RANGE

Connect the output leads of the PIE 311Plus to the inputs of the device being calibrated.

Instantly output your SPAN and ZERO output settings by moving the EZ-CHECK switch between HI 
and LO. You may also select any third output setting (such as mid-range) using the SET position on 
the EZ-CHECK switch. The output is adjusted in 0.01 Ω increments by turning the knob 

e

. Press 

and turn the knob for faster dialing with 1.00 Ω increments.

Voltage Receiver Input

Controller

Transmitter

Computer

Logger

DCS

        2                1   

 

     

           

  

OFF

OVERLOAD

Double Click 

Menu

HI

SET

LO

SOURCE

MAX
READ

MIN

READ

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Diagnostic RTD &

Milliamp Calibrator

2, 3, 4 Wire Detection

Loop Leak Detect ∙ Transmitter Supply

+

-

mV

40.000

All RTD & Ohms Connections

Voltage Output Signal

Controller

Transmitter

Power Supply

        2                1   

 

     

           

  

OFF

OVERLOAD

Double Click 

Menu

HI

SET

LO

SOURCE

MAX
READ

MIN

READ

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Diagnostic RTD &

Milliamp Calibrator

2, 3, 4 Wire Detection

Loop Leak Detect ∙ Transmitter Supply

+

-

mV

42.345

Read Ohms

Choose this function to measure from 0.00 to 4010.0 ohms.

Move the power switch 

to READ then Double Click the EZ-DIAL knob to get into the Menu. Turn 

the knob to scroll through the settings and press the knob to make your selection. Select 

OHMS

 for 

the 

FUNCTION

 and 

400Ω 

or

 4000Ω

 for the 

RANGE

Connect the input leads of the PIE 311Plus to the resistance being measured.

Signals above the maximum scale are limited by protection circuitry with “OVER RANGE” flashed on 
the display and the red OVERLOAD LED lit.

The PIE 311Plus measures the input signal and constantly updates the display with the current 
reading. Move the EZ-CHECK switch 

to MAX to see the highest reading and to MIN to see the 

lowest reading. Press and hold the knob 

e

 to clear the MAX and MIN readings.

Содержание 311Plus

Страница 1: ...for simple hookups Carry it without worry it comes protected with a rubber boot and rugged low profile switches Easy to operate even in dark areas with the backlit display Calibrate directly in temper...

Страница 2: ...ectronics 82 East Main Street Suite 3 14 Webster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Magnetic Hanging Strap Optional Deluxe Hands Free Carrying Case Included Evo...

Страница 3: ...and turn the EZ DIAL knob for faster dialing Press and hold the knob for two seconds to store desired EZ Check HI LO points in SOURCE mode Continue to press and hold the knob for three more seconds to...

Страница 4: ...0 a 3850 Pt 200 a 3850 Pt 500 a 3850 Pt 1000 a 3850 Pt 100 a 3902 Pt 100 a 3916 Pt 100 a 3926 Cu 10 a 4274 Cu 50 a 4280 Ni 120 a 6720 Pt 10 a 3850 Pt 50 a 3850 Read mA Power Measure Transmitters Leak...

Страница 5: ...EXIT 4 4 AUTO OFF ON OFF BACKLIGHT ON OFF STEPS RAMP 2 3 5 11 RAMP STEP RAMP TIME 5 6 7 8 9 10 15 20 25 30 60 Double Click Menu FEATURES Double Click Menu mA DISPLAY MODE pressing the knob will cycle...

Страница 6: ...he EZ CHECK Switch into the HI or LO position then press and hold the e DIAL KNOB for 6 seconds the word STORE will appear on the display after 3 seconds and continue to press the DIAL KNOB until the...

Страница 7: ...nections are made at the top of the calibrator where the jacks are numbered for ease of use SOURCE Plug in the 2 3 or 4 wires to match the connection on the instrument being calibrated Connections REA...

Страница 8: ...made to a 4 wire sensor and the 311Plus indicates that only Wires 1 2 4 are connected There may be a loose connection or a break in wire 3 somewhere between the sensor and the 311Plus Only a calibrato...

Страница 9: ...calibrated and connect the PIE 311Plus to the inputs of the device using 2 3 or 4 wires The sensor current generated by the instrument will be indicated on the display followed by the word FIXED or PU...

Страница 10: ...Detect Transmitter Supply C 3 2 1 4 100 0 OFF OVERLOAD Double Click Menu HI SET LO SOURCE MAX READ MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4...

Страница 11: ...AD MIN READ Push Hold to Store Step Push Turn for Fast Dialing Diagnostic RTD Milliamp Calibrator 2 3 4 Wire Detection Loop Leak Detect Transmitter Supply mV 40 000 All RTD Ohms Connections Voltage Ou...

Страница 12: ...such as mid range using the SET position on the EZ CHECK switch The output is adjusted in 0 001 mA 0 01 increments by turning the knob e Press and turn the knob for faster dialing with 0 100 mA 1 00 i...

Страница 13: ...and disconnect the PIE 311Plus Signals above 24 mA are current limited by protection circuitry with OVERRANGE flashed on the display and the red OVERLOAD LED lit To Sensor Typical 2 Wire Transmitter D...

Страница 14: ...oose READ if your would like the mA display in percent of 4 20 milliamps PWRM is POWER MEASURE which uses the internal loop supply of the 311Plus to power up the transmitter while indicating the curre...

Страница 15: ...ster NY 14580 Tel 585 872 9350 Fax 585 872 2638 sales piecal com www piecal com Operating Instructions Typical 2 Wire Transmitter IN REF OUT 311 2 F HART 5 6 PWRM 12 523 mA Pt 100 3850 00 51 mA LEAK E...

Страница 16: ...PIE 311Plus supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter or loop The current passed by the transmitter will be accurately displayed by the 311Plus along with an indication of leak...

Страница 17: ...inst misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas Read Ohms RTD 3 Wire 4 Wire Ac...

Страница 18: ...003 mA Voltage burden 2V at 24 mA Overload Current limit protection 25 mA nominal Source mA Power Measure Two Wire Transmitters PWRM LEAK Ranges and Resolution 0 000 to 24 000 mA or 25 00 to 125 00 of...

Страница 19: ...0 842 0 to 1256 0 1256 0 to 1562 0 0 1 0 2 0 3 0 4 0 5 Pt 500 Ohm DIN IEC JIS 1989 ITS 90 1 3850 0 00385 200 0 to 90 0 120 0 to 180 0 180 0 to 450 0 450 0 to 680 0 680 0 to 850 0 0 1 0 1 0 2 0 2 0 3 3...

Страница 20: ...other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Inform...

Отзывы: