5 Installation
U651 Rotation Stage
MP136E
Version: 1.3.0
19
Figure 6: U-651: Holes in the base body for M6 screws (white arrows) or M4 screws (black arrows)
Requirements
You have read and understood the general notes on installation (p. 15).
You have provided a suitable underlying surface (see "Dimensions" (p. 42) for the
required position of the holes for the screws).
−
Four M6 or M4 holes are provided with a thread depth of at least 13 mm.
−
The surface flatness is ≤10
µm.
−
For applications with large temperature changes: The surface should have the same
thermal expansion properties as the U-651 (e.g., surface made of aluminum).
You have accounted for the space required to route cables without bending and
according to regulations.
The U-651 is
not
connected to the controller.
Tools and accessories
Mounting accessories in the scope of delivery (p. 10):
−
Four M6 screws
−
Hex key
Alternative (not in the scope of delivery):
−
Four M4 screws of suitable length
−
Suitable screwdriver