Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
P-726.1CD
Unit
Tolerance
Miscellaneous
Operating temperature range
-20 to 80
°C
Material
Aluminum, steel
Objective thread
M32
Mass
575
g
±5 %
Cable length
1.5
m
±10 mm
Sensor/voltage connection
D-sub 7W2 (m)
Recommended electronics
E-505, E-621, E-625, E-665,
E-709, E-754
The resolution of the system is limited only by the noise of the amplifier and the measuring
technology because PI piezo nanopositioning systems are free of friction.
All specifications based on room temperature (22 °C ±3 °C).
Ask about customized versions.
Maximum Ratings
P-726 objective scanners are designed for the following operating data:
Maximum operating voltage Maximum operating
frequency (unloaded)
1
Maximum power
consumption
2
-20 to 120 V
100 Hz
20 W
1
To ensure stable operation, the maximum operating frequency is defined as approximately
1/3 of the mechanical resonant frequency.
2
The heat generated by the piezo actuator during dynamic operation limits the value for
maximum power consumption.
Details can be found online: