User Manual
P726T0002, valid for P-726.1CD
CBo, 10/18/2019
Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Page 9 / 17
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email
Installation
Connecting the P-726 to a Protective Earth Conductor
The P-726 has no separate protective earth connection. The system into which the P-726 is
integrated (e.g., surrounding mechanical system) must be connected to a protective earth
conductor (required cross-
sectional area of the cable ≥0.75 mm
2
).
When you connect the P-726 to a protective earth conductor:
1.
Make sure that the contact resistance is
<0.1 Ω at 25 A
at all connection points relevant
for mounting the protective earth conductor.
2.
Observe the applicable standards for mounting the protective earth conductor.
Affixing the P-726 to the Microscope
Requirements
You have read and understood the safety precautions (p. 5).
Tools and accessories
P726T0001 technical note for PIFOC QuickLock Thread Option for P-726 (p. 8)
QuickLock thread adapter (p. 8) (not included)
Suitable tools
Affixing the P-726 to the microscope
Install the P-726 as described in the P726T0001 technical note for PIFOC QuickLock
thread option (p. 7).
Installing the Objective
Correct:
Figure 2: Center of load close to aperture