10 Technical Data
P-5x7/P-5x8 Positioners
PZ82E
Version: 3.2.0
41
Model
Maximum operating
voltage
Maximum operating
frequency (unloaded)
1
Maximum power
consumption
2
P-558.ZCD
P-558.ZCL
P-558.TCD
-20 to 120 V
190 Hz
22 W
P-518.ZCD
P-518.ZCL
P-518.TCD
-20 to 120 V
167 Hz
28 W
P-528.ZCD
P-528.ZCL
P-528.TCD
-20 to 120 V
117 Hz
34 W
1
To ensure stable operation, the maximum operating frequency has been defined as around
one third of the mechanical resonant frequency.
2
The heat that is generated by the piezo actuator during dynamic operation limits the value for
maximum power consumption.
Details can be found at the following website:
https://www.physikinstrumente.com/en/technology/piezo-technology/properties-piezo-
actuators/electrical-operation/
10.2
Ambient Conditions and Classifications
Pay attention to the following ambient conditions and classifications for the P-5x7/P-5x8:
Area of application
For indoor use only
Maximum altitude
2000 m
Air pressure
1100 hPa to 700 hPa
Relative humidity
Highest relative humidity 80 % for temperatures up to 31 °C
Decreasing linearly to 50 % relative humidity at 40 °C
Operating temperature
-20 °C to 80 °C
Storage temperature
-20 °C to 80 °C
Transport temperature
-25 °C to 85 °C
Overvoltage category
II
Protection class
I
Degree of pollution
1
Degree of protection
according to IEC 60529
IP20