Overview
1. Connector for the protective earth conductor on the L-505.01A200F, indicated by the
protective earth symbol
2. Safety washer
3. Flat washer
4. Protective earth conductor lug
5. Screw, ISO 7045 M4×8
Tools and Accessories
■
Suitable protective earth conductor:
■
Cable cross section
≥
0.75 mm
2
■
Contact resistance <0.1
Ω
at 25 A at all connection points relevant for attaching the
protective earth conductor
■
M4 screw set for protective earth, in the
scope of delivery of the L-505.01A200F (p. 10)
■
Suitable screwdriver
Requirements
✓
You have read and understood the
general safety instructions (p. 8)
.
Information
►
Observe the applicable standards for connecting the protective earth conductor.
Connecting the L-505.01A200F to the Protective Earth Conductor
1. If necessary, attach a suitable cable lug to the protective earth conductor.
2. Attach the cable lug of the protective earth conductor to the protective earth connection
as shown in the illustration of the L-505.01A200F.
3. Tighten the screw with a torque of 1.2 Nm to 1.5 Nm.
4. Make sure that the contact resistance is <0.1
Ω
at 25 A at all connection points relevant for
attaching the protective earth conductor.
6.3
Setting Up a Multi-Axis System
The L-505.01A200F can be used in XY systems.
Designations in these instructions:
■
Lower positioner: Forms the basis of the multi-axis system, is attached to an underlying
surface
■
Upper positioner: Forms the upper axis of the multi-axis system, is attached to the lower
positioner rotated by 90°
6 INSTALLATION
L505M0011EN
‒
10/22/2018
15
M O T I O N | P O S I T I O N I N G