3.20.
IRIG Time Code (External Time Synchronization)
The system supports IRIG-B input and can add an IRIG code to each recorded frame. The sample
timing for the IRIG code is once each frame.
The recorded IRIG code is displayed with the PFV.
IRIG Code Input Specification
Connector
BNC
Code Format
IRIG-B (122) Analog
Amplitude
3.0Vp-p min,8.0Vp-p max
Mark to space ratio
3:1 to 6:1
Typical modulated carrier signal ratio
10:1
•
Please refer to “3.8 Other Settings” of “Photron FASTCAM Viewer User’s Manual” for the details of
the setup.
•
IRIG Time Code is used when synchronizing a camera with external equipment in time.
It is a convenient function when apparatus is physically separated.
•
When the IRIG code is being input, the IRIG code is displayed in white, and is displayed to the left.
•
The IRIG offset time is also displayed below it. When the IRIG code is not being input, the IRIG
code is displayed in grey. At that time, the counter is the camera’s internal counter and it continues
to count.
59
FASTCAM Mini WX50/100 Hardware Manual
Содержание FASTCAM Mini WX100
Страница 1: ...F FA AS ST TC CA AM M M Mi in ni i W WX X5 50 0 1 10 00 0...
Страница 12: ......
Страница 26: ...Chapter 1 Overview 14...
Страница 40: ...Chapter 2 Setup 28...
Страница 73: ...61 FASTCAM Mini WX50 100 Hardware Manual...
Страница 76: ...Chapter 5 Product Specifications 5 1 Specifications 5 2 Dimensions 64...
Страница 80: ...Chapter 5 Product Specifications 68...
Страница 94: ...Chapter 5 Product Specifications 5 2 2 AC Power Supply Unit mm 82...
Страница 95: ...83...
Страница 96: ...Chapter 6 Warranty Chapter 6 Warranty 6 1 About the Warranty 84...
Страница 98: ...Chapter 7 Contacting Photron Chapter 7 Contacting Photron 7 1 Contact Information 86...