Chapter 6. Specifications

29

Input/Output Port

Pinout

The input/output port provides information about trigger function and shutter 
status. All inputs are pulled up to +5V DC through 10k ohm resistors and 
filtered with 2200pf ceramic capacitors. Outputs are driven by a 74F374 latch. 

The numbers on the trigger connector diagram below correspond with the 
numbers given to the definition of each of the pins. The I/O connector is a 
female, D-subminiature 9-pin connector.

Input/Output Port with Pin Numbers

1

SHUTTER OPEN OUTPUT
Active high. A high level on this output indicates that the shutter is 
completely open. The output is low when the shutter is closed or in 
motion.

2

SHUTTER MOVING OUTPUT
Active high. A high level on this output indicates that the shutter is 
opening or closing. The output is low when the shutter is completely open 
or completely closed. The length of time this signal is held high is specified 
in PVCAM

®

 or by the Shutter Open Delay and Shutter Close Delay 

functions in the software.

3

CAMERA EXPOSING OUTPUT
Active high. A high level on this output indicates that the camera is 
exposing (integrating). This output does not include shutter open and 
close delay times.

4

FLASH 
This pin is a TTL-level output that can be controlled through software by 
using the FLASH command.

5

FACTORY SETUP 
This pin is used for factory setup only and is not to be utilized by the user. 
Driving this pin will disable the camera and cause a PVCAM error.

6

FILTERED TRIGGER INPUT
This input uses a TTL voltage level to trigger the camera. The input is 
pulled low to initiate a trigger (default). If nothing is connected to the 
input, it is pulled high to prevent the system from getting false triggers. To 
change the state of this input see TRIGGER INVERT INPUT (8). (The 

Содержание SenSys

Страница 1: ...SenSys User Manual...

Страница 2: ...SenSys User Manual...

Страница 3: ...f Nikon Corporation Other brand and product names are the trademarks or registered trademarks of their respective owners and manufacturers The information in this publication is believed to be accurat...

Страница 4: ...xcept for removing the shutter cover this warranty is voided if the user opens the camera head Metachrome II Warranty Photometrics Ltd warrants that the Metachrome II coating will maintain 95 of initi...

Страница 5: ...thout prior authorization by Photometrics The buyer shall prepay shipping charges to Photometrics and Photometrics shall pay shipping charges to return the product to the buyer for products returned t...

Страница 6: ...on of the laws of Member States relating to Electromagnetic Compatibility and Product Safety This declaration is based upon compliance of the product to the following standards EN 55022 A CISPR 22 RF...

Страница 7: ......

Страница 8: ...s and Other Instruments 5 Connecting the AIA Cable 5 Connecting the Power Brick 6 Chapter 3 Configuration Options Lens Mount Adapters 7 C Mount Assembly 8 F Mount Assembly 9 Tripod Camera Stand 10 Cha...

Страница 9: ...19 Camera Shutter Needs Replacing 20 Camera Does Not Focus 22 Chapter 6 Specifications Camera 23 F Mount Camera 24 C Mount Camera 25 F Mount Adapter 26 C Mount Adapter 27 Power Brick 28 Input Output P...

Страница 10: ...rade charge coupled device CCD SenSys System Components All SenSys systems consist of standard hardware and software as well as the appropriate interface hardware discussed in the Software Guide for y...

Страница 11: ...can be operated in any orientation therefore it can be used on a tripod camera stand or mounted on a variety of instruments Storage Requirements Store the SenSys camera system in its original contain...

Страница 12: ...hotometrics Customer Service If you have any questions contact Photometrics Customer Service When you call please have your Photometrics job number or equipment serial numbers available Phone 520 889...

Страница 13: ...4 SenSys User Manual...

Страница 14: ...r connect your own trigger equipment to the 9 pin female input output I O trigger port located on the back of the camera If you are using high voltage equipment such as an arc lamp with your camera sy...

Страница 15: ...era see the diagram on page 14 Secure by tightening the connector screws Do not extend this cable 2 With the power brick switched off and the power cord unplugged from an electrical outlet connect the...

Страница 16: ...vable F mount adapter or C mount adapter installed The F mount adapter allows the camera to be attached to lenses and scientific instruments that are compatible with a Nikon bayonet mount The C mount...

Страница 17: ...mount adapter onto the lens or instrument adapter Lock in place using the 050 hex wrench provided 3 Thread the C mount adapter into the camera In certain cases it may be easier to thread the camera on...

Страница 18: ...e a lens simply attach your instrument s F mount adapter to the camera or attach the camera to the instrument s F mount If your instrument requires an F mount adapter contact your instrument supplier...

Страница 19: ...so that the maximum length of the tripod mounting bolt is exposed 2 Align the camera on the tripod mounting bolt using one of the four mounting holes on the camera body 3 Thread the tripod mounting bo...

Страница 20: ...apter and Nikon Lens CCD When you order a SenSys camera you choose from a range of CCDs that differ in size and grade All SenSys CCDs are scientific grade grades with fewer defects than commercial gra...

Страница 21: ...on than crown glass The antireflective coating increases the light transmitted through the window from 92 to 99 further decreasing light loss Optional infrared blocking IR and sapphire windows are ava...

Страница 22: ...ved by the ADC If the ADC receives a high quality signal the 12 bit ADC digitizes the analog signal into 12 bit data With a lesser quality signal the ADC produces data with a lower effective bit depth...

Страница 23: ...specifications for the port are located in Power Port Pinout on page 31 Camera Back Status Lights The back of the camera has two status lights Red light on Power on Green light on CCD reached operati...

Страница 24: ...on and off 0 by a switch on the brick More detailed specifications are available in Chapter 6 Specifications Caution Connecting or removing a live power cable can damage the camera s electronic compo...

Страница 25: ...he SenSys F and C mount adapters thread into the camera at a pitch of 32 threads inch When the camera is focused two 6 32 inch socket set screws two 4 40 inch socket set screws for the C mount secure...

Страница 26: ...e F mount adapter is a standard Nikon bayonet mount with a standard F mount flange focal distance and additional travel distance to allow parfocal adjustment with a variety of instruments For specific...

Страница 27: ...18 SenSys User Manual...

Страница 28: ...Camera Does Not Respond to Light If your camera has no response to light a faulty camera shutter may be causing the problem To test for a malfunctioning shutter 1 Turn on the power to the camera 2 As...

Страница 29: ...hex wrench remove the four 4 40 socket head screws from the shutter cover Caution Make sure to remove the shutter cover and not the camera back Removing the camera back will expose the CCD and void yo...

Страница 30: ...rd Then unplug the shutter power connector from the shutter board 4 Using a standard screwdriver remove the four 2 56 slotted binding head screws that secure the shutter to the camera Lift the shutter...

Страница 31: ...mount adapter is installed on your camera and the mount is attached directly to an instrument See the instructions for focusing the camera in F Mount Assembly on page 9 A C mount adapter is installed...

Страница 32: ...ge plane CCD F mount Distance C mount Distance 0400 and 1600 1 05 in 26 7 mm 63 in 16 0 mm 1400 1 07 in 27 2 mm 65 in 16 6 mm Distance from top of window to CCD image plane CCD F mount Distance C moun...

Страница 33: ...24 SenSys User Manual F Mount Camera F mount Camera SideView F mount CameraTopView 3 00 in 76 2 mm 7 15 in 181 6 mm 4 48 in 113 8 mm 3 00 in 76 2 mm 4 48 in 113 8 mm 3 00 in 76 2 mm...

Страница 34: ...Chapter 6 Specifications 25 C Mount Camera C mount Camera SideView C mount CameraTopView 7 15 in 181 6 mm 2 60 in 66 0 mm 4 48 in 113 8 mm 2 60 in 66 0 mm...

Страница 35: ...Top View of F mount 2 35 in dia 59 7 mm 2 50 in dia 63 5 mm 1 05 in 26 7 mm 1 72 in 43 7 mm minimum distance 2 50 in dia 63 5 mm 1 93 in dia 49 0 mm 67 in 17 0 mm 1 02 in 25 9 mm Threads at 32 threads...

Страница 36: ...dia 38 1 mm 96 in dia 24 5 mm 1 60 in dia 40 6 mm Side View of C mount unmounted Side View of C mount mounted for KAF 1400 Side View of C mount mounted for KAF 0400 and KAF 1600 Top View of C mount 1...

Страница 37: ...28 SenSys User Manual Power Brick The power supply is a 5V DC and 15V DC brick with 100 240V AC input at 50 60 Hz The maximum power output is 55 W 6 97 in 177 0 mm 3 74 in 95 0 mm 2 25 in 57 1 mm...

Страница 38: ...ng The output is low when the shutter is completely open or completely closed The length of time this signal is held high is specified in PVCAM or by the Shutter Open Delay and Shutter Close Delay fun...

Страница 39: ...enabled The inputs are internally pulled up therefore it is recommended to drive them with an open collector driver 8 TRIGGER INVERT INPUT Active low A low on this input inverts the state of FILTERED...

Страница 40: ...Chapter 6 Specifications 31 Power Port Pinout Power Port Pinout Pin Signal Name 1 5V DC 2 5V DC 3 Ground 4 Ground 5 Ground 6 15V DC 7 Ground 8 15V DC 9 Ground...

Страница 41: ...4 VD13 27 50 VD2 5 VD12 28 51 6 VD11 29 PIX 52 7 VD10 30 53 VD1 8 VD9 31 54 VD0 9 VD8 32 55 10 VD7 33 56 TX 11 VD6 34 Ground 57 RX 12 Ground 35 Ground 58 13 VD5 36 VD15 59 FEN 14 VD4 37 VD14 60 LEN 1...

Страница 42: ...Chapter 6 Specifications 33 CCD Orientation Camera FrontView KAF 0400 and KAF 1600 CCD Orientation Camera FrontView KAF 1400 CCD Orientation Serial Parallel 0 0 Parallel Serial 0 0...

Страница 43: ...Gain Setting Relative Gain Typical Noise System Gain Typical Maximum ADC Signal High Signal to Noise 1 1 2x 26e 40e ADU 160Ke binned full well High Dynamic Range 2 1x 19e 20e ADU 80Ke single pixel ful...

Страница 44: ...ion Mode Gain Setting Relative Gain Typical Noise System Gain Typical Maximum ADC Signal High Signal to Noise 1 1 2x 20e 20e ADU 80Ke binned full well High Dynamic Range 2 1x 17e 10e ADU 40Ke single p...

Страница 45: ...Mode Gain Setting Relative Gain Typical Noise System Gain Typical Maximum ADC Signal High Signal to Noise 1 1 2x 26e 40e ADU 160Ke binned full well High Dynamic Range 2 1x 19e 20e ADU 80Ke single pix...

Страница 46: ...enSys camera 2 F F mount adapter 9 16 description 17 installation 9 removal 9 role in focusing camera 9 specifications 26 G gain available settings 13 H hardware installation 5 6 hardware specificatio...

Страница 47: ...13 mounting on tripod 10 precautions 2 primary point digitization PPD 13 readout rate 13 repair 3 specifications 23 status lights 14 storage requirements 2 system components 1 turning on 6 window 12...

Страница 48: ...57 054 001 Rev A...

Отзывы: