20
SenSys User Manual
Camera Shutter Needs
Replacing
If your camera's shutter signal is working but the shutter needs to be replaced:
1. Using a 3/32” hex wrench, remove the four #4-40 socket head screws
from the shutter cover.
Caution: Make sure to remove the shutter cover and not the camera back. Removing
the camera back will expose the CCD and void your warranty.
2. Using a #1 Phillips screwdriver, remove the two #4-40 pan head
Phillips screws that attach the shutter board.
Shutter Cover
Shutter Board