6-8
Operation
Figure 6-1: PPV waveform – Max V limit
When the Max P (PPV maximum pressure limit) is reached, pressure is limited
but the breath is not terminated, and a message is displayed. After the limit is
reached in three consecutive breaths, the audible alarm sounds. A PPV
waveform with Max P is shown in Figure 6-2.
Figure 6-2: PPV waveform – Max P limit
Frequent annunciation of one or both alarms typically indicates improved
patient status. It may, however, indicate that the patient is more actively
breathing, possibly due to agitation or a change in the patient’s level of
sedation. It may also indicate an increase in leakage.
The V
T
(estimated exhaled tidal volume) measurement may remain below the
set Max V limit even though the inspired volume exceeds Max V. This results
from variable leakage, which reduces the exhaled volume in relation to the
inspired volume.
Vol
u
me
Fl
o
w
Max V
V
T
Max V
Inspiration terminated
Sudden increase
in flow
Time
Pressure
EPAP
Max P
Sudden increase
in target pressure
Pressure limited
Exhalation cycled by patient
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......