Alarms, messages, and troubleshooting
9-15
Check Vent: Ma-
chine Pressure Sen-
sor Calibration Data
Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(7)
Yes
No
No
Check Vent: Ma-
chine Pressure Sen-
sor Range Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(11)
Yes
No
No
Check Vent: Motor
Control PCBA ADC
Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(28)
Yes
No
No
Check Vent: O
2
Flow
Sensor Calibration
Data Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(14)
Yes
No
No
Check Vent: O
2
Pres-
sure Sensor Calibra-
tion Data Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(15)
Yes
No
No
Check Vent: O
2
Sup-
ply Pressure Sensor
Range Error
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(17)
Yes
No
No
Check Vent: OVP
Circuit Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(38)
Yes
No
No
Check Vent: Oxygen
Device Failed
Continues to ventilate
with air only
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(16)
Yes
No
No
Check Vent: Primary
Alarm Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(4)
Yes
No
No
Check Vent: Pro-
gram CRC Test Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(1)
Yes
No
No
Check Vent: Proximal
Pressure Sensor
Auto-Zero Failed
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(10)
Yes
No
No
Check Vent: Proximal
Pressure Sensor Cali-
bration Data Error
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(8)
Yes
No
No
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......