11-2
Technical specifications
Ramp Time
OFF, 5 to 45 min
5 min
± 1 sec
OFF
Rate (Respiratory Rate)
4 to 60 BPM
1 BPM
± 1 BPM
4 BPM
Rise (Rise Time)
1 to 5
1
N/A
3
V
T
(AVAPS Target Tidal
Volume)
200 to 2000 mL BTPS
5 mL
± 15%
500 mL
Table 11-1: Control setting ranges, resolutions, and accuracies (continued)
Parameter
Range
Resolution
Performance Accuracy
Factory default
Table 11-2: Controls active in Respironics V60/V60 Plus ventilation modes
CPAP
S/T
PCV
AVAPS
PPV
Timing
Rate
Rate
*
I-Time
I-Time*
Baseline pressure
CPAP
EPAP
Inspiratory pres-
sure
IPAP
Max P
Max P
Min P
IPAP*
Rise Time
Rise
Rise*
O
2
O
2
Volume
V
T
Max V
Ramp feature
Ramp Time
Mode-specific
C-Flex
PPV %
Max E
Max R
* Used in backup only
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......