Regulatory compliance
D-5
Recommended separation distances between
portable and mobile RF communications equipment and the V60/V60 Plus Ventilator
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
outside ISM bands
150 kHz to 80 MHz
in ISM bands
80 MHz to 800 MHz
800 MHz to 2.7 GHz
0.01
0.12
0.20
0.40
0.77
0.1
0.37
0.63
1.26
2.42
1
1.17
2.00
4.00
7.67
10
3.69
6.32
12.65
24.24
100
11.67
20.00
40.00
76.67
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in me-
ters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maxi-
mum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795
MHz;13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
NOTE 3: An additional factor of 10/3 has been incorporated into the formulae used in calculating the recommended
separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency
range 80 MHz to 2.7 GHz to decrease the likelihood that mobile/portable communications equipment could cause
interference if it is inadvertently brought into patient areas.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE 5: The minimum separation distance for RF communication equipment operating within the following fre-
quency bands is 0.3 m:
• 380 –390 MHz (TETRA 400)
• 430 – 470 MHz (GMRS 460, FRS 460)
• 704 – 787 MHz (LTE Band 13, 17)
• 800 – 960 MHz (GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5)
• 1 700 –1 990 MHz (GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS)
• 2 400 – 2 570 MHz (Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7)
• 5 100 – 5 800 MHz (WLAN 802.11 a/n)
d
3.5
V
1
-------
P
=
d
12
V
2
-------
P
=
d
12
E
1
-------
P
=
d
23
E
2
-------
P
=
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......