9-14
Alarms, messages, and troubleshooting
Check Vent: Blower
Stalled
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(2)
Yes
No
No
Check Vent: Blower
Temperature High
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(33)
Yes
No
No
Check Vent: Cooling
Fan Speed Error
Overheating of ventila-
tor possible
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(36)
Yes
No
No
Check Vent: CPU
PCBA ADC Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(29)
Yes
No
No
Check Vent: Data Ac-
quisition PCBA ADC
Failed
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(27)
Yes
No
No
Check Vent: Flash
File System Error
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(37)
Yes
No
No
Check Vent: Internal
Temperature High
CPU
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(30)
Yes
No
No
Check Vent: Internal
Temperature High
Daq
Technical failure
Check the patient. Check for
causes of overheating, such
as high room temperature,
blocked vents, clogged air
inlet filter, or nonfunctional
fan. Provide alternative ven-
tilation. Have the ventilator
serviced.
High
(31)
Yes
No
No
Check Vent: Internal
Temperature High
Mtr
Technical failure
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(32)
Yes
No
No
Check Vent: Ma-
chine Pressure Sen-
sor Auto-Zero Failed
Proximal pressure is
not measured. Pres-
sure-related alarms are
compromised.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
High
(9)
Yes
No
No
Table 9-4: Check Vent alarm messages: summary and troubleshooting (continued)
Message
Description
Corrective Action
Priority
type (ID)
Manually
resettable
Autores
ettable
Silence
able
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......