EN/IEC 60601
Test Level
±1 kV line(s)
to line(s)±2 kV
line(s) to earth
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle.
40% U
T
(60% dip
in U
T
) for 5
cycles.
70% U
T
(30% dip
in U
T
) for 25
cycles
5% U
T
(>95% dip
in U
T
) for 5 sec.
Compliance
Level
±1 kV line(s)
to line(s)±2 kV
line(s) to earth
Immunity Tes
t
Surge EN/IEC
61000-4-5
(only for devices with battery
charger power supply
-700
0
MHB & 7100MHB)
3A/m
3A/m
Power Frequency
50/60Hz Magnetic
Field EN/IEC
61000-4-8
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines EN/
IEC 61000-4-11
(only for devices with
battery charger power
supply -700
0
MHB &
7100MHB)
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle.
40% U
T
(60% dip
in U
T
) for 5
cycles.
70% U
T
(30% dip
in U
T
) for 25
cycles
5% U
T
(>95% dip
in U
T
) for 5 sec
.
Electromagnetic
Environment – Guidanc
e
NOTE : U
T
is the a.c. mains voltage prior to application of the test level.
Electrical
power quality
should be that of a typical
commercial or hospital
environment.
During
charging, i
f the the user of the
GoSafe
(
7000MHB/
71
0
0MHB
)
requires continued operation
during power mains
interruptions, it is
recommended that the
GoSafe device power adapter
is
powered from an
uninterrupted power supply
or a battery.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.