Maintaining, Testing, and Troubleshooting Your HeartStart FR2+
4-15
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
4
Status Indicator:
SOLID
RED
X
Audio signal:
NONE
•
The battery is missing or completely
depleted.
•
The training & administration pack is
being used in the administration
function (the solid red X is normal in
this case) or has been left in the FR2+
by mistake.
•
A self-test detected a failure.
•
Install a
new
M3863A FR2
standard battery or a charged
M3848A FR2+ rechargeable
battery in the FR2+ and perform
the battery insertion test (BIT).
•
Remove the training &
administration pack and install a
battery.
•
Remove and reinstall the battery
and perform the battery insertion
self-test. If it fails, install a
new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
Status Indicator:
SOLID
RED
X
Audio signal:
CHIRPING
•
The training & administration
pack is being used in the
ADMINISTRATION
function
and more than 10 minutes have
passed without user interaction
(button press or pads change).
•
The training & administration pack is
being used in the TRAINING function
and more than 30 minutes have
passed without user interaction
(button press or pads change).
•
To continue using the training &
administration pack, press any
button (except On/Off).
•
To return the FR2+ to standby
mode, remove the Pack and install
a battery.
Status Indicator:
NONE
The FR2+ has been physically damaged.
•
Check for visible damage. Do not
use the FR2+ if it appears to be
damaged.
•
Remove and reinstall the battery to
perform the battery insertion
self-test. If it fails, install a
new
M3863A FR2 standard battery or
a charged M3848A FR2+
rechargeable battery and repeat
the test. If it fails again, do not use
the FR2+.
symptom
possible cause
recommended action
Содержание HeartStart FR2+
Страница 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Страница 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Страница 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...