After Using the HeartStart FR2+
After each use of the FR2+, perform the maintenance tasks described in the
table above, as well as the following post-use checks before returning the
FR2+ to service:
Check the operation of the FR2+
by removing and reinstalling the
battery and running the battery insertion selftest.
N O T E
:
Perform also
when replacing expired defibrillator pads.
Check the outside of the FR2+
and the connector socket for signs of
dirt or contamination. If the FR2+ is dirty or contaminated, clean it
according to the guidelines provided in this manual.
Check the data card
if one has been used. If the data card has been
used to record incident data, remove and replace it with a blank data
card. Deliver the recorded data card to appropriate personnel according
to your local guidelines and medical protocol.
daily
monthly
maintenance task/response
Check the Status Indicator.
If you see the flashing black hourglass: The FR2+ is
ready to use. No action required.
If you see anything other than a flashing black
hourglass, remove and reinstall the battery to run the
selftest.
• If the selftest passes and the Status Indicator
shows the flashing black hourglass, the FR2+ is
ready to use.
• If the selftest fails, install a new battery and run the
selftest. If the selftest passes, the FR2+ is ready to
use. If the selftest fails, contact Philips Medical
Systems.
Check supplies, accessories, and spares
for
damage and expiration dating.
Do not use damaged or expired accessories. Replace
them immediately.
If a
LOW
BATTERY
or
REPLACE
BATTERY
message is
displayed: Replace the battery and run the selftests.
DO
NOT
ATTEMPT
TO
CHARGE
THE
M3863A
FR2
STANDARD
BATTERY
. It is not rechargeable. The
M3848A FR2+ battery is rechargeable. Recharge it,
using the M3849A Charger, for the FR2+
rechargeable battery only.
Check the outside of the FR2+
and the connector
socket for cracks or other signs of damage.
If you see signs of damage: Contact Philips Medical
Systems for technical support.
Содержание HeartStart FR2+
Страница 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Страница 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Страница 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...