Energy
Using adult defibrillator pads: 150 J nominal into a 50 ohm load.
Using infant/child reduced-energy defibrillator pads: 50 J nominal into a 50 ohm
load. Sample pediatric energy doses:
age
energy dose
newborn
14 J/kg
1 year
5 J/kg
2
−
3 years
4 J/kg
4
−
5 years
3 J/kg
6
−
8 years
2 J/kg
Doses indicated are based on CDC growth charts for the 50th percentile weights
for boys.*
*
National Center for Health Statistics in collaboration with the National Center for
Chronic Disease Prevention and Health Promotion.
CDC growth charts:
weight-for-age percentiles, revised and corrected
November 28, 2000. Atlanta,
GA: Centers for Disease Control and Prevention © 2000.
Charge Control
Controlled by Patient Analysis System for automated operation. Can be
programmed for manual initiation using advanced mode of the M3860A.
Charge Time
from “Shock Advised”
< 10 seconds typical, including confirming analysis. Charge time increases near
end of battery service life.
Shock-to-Shock Cycle Time
< 20 seconds typical, including analysis, in AED mode.
“Charge Complete”
Indicator
Shock button flashes, audio tone sounds.
Disarm (AED mode)
Once charged, the HeartStart FR2+ will disarm if:
• patient’s heart rhythm changes to non-shockable rhythm,
OR
• a shock is not delivered within 30 seconds after the FR2+ is armed,
OR
• the PAUSE button (if enabled) is pressed,
OR
• the On/Off button is pressed to turn off the FR2+,
OR
• the defibrillator pads are removed from the patient or the pads connector is
disconnected from the FR2+.
category
nominal specifications
Содержание HeartStart FR2+
Страница 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Страница 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Страница 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...