C A U T I O N
— conditions, hazards, or unsafe practices that can
result in minor personal injury, damage to the HeartStart FR2+, or
loss of data stored in the device.
THESE SAFETY CONSIDERATIONS ARE DIVIDED INTO FOUR
GROUPS: SAFETY CONCERNS ABOUT THE HEARTSTART FR2+ IN
GENERAL USE, DEFIBRILLATION, MONITORING, AND MAINTENANCE
ACTIVITIES.
The dangers, warnings, and cautions listed in the following tables apply to
both the model M3860A and the model M3861A HeartStart FR2+, unless
otherwise noted.
General dangers, warnings, and cautions
safety level
possible shock or fire hazard, or explosion
DANGER
THERE IS A POSSIBILITY OF EXPLOSION IF THE HEARTSTART
FR2+ IS USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS
OR CONCENTRATED OXYGEN.
DANGER
THE HEARTSTART FR2+ HAS NOT BEEN EVALUATED OR
APPROVED FOR USE IN HAZARDOUS LOCATIONS AS DEFINED IN
THE NATIONAL ELECTRICAL CODE (ARTICLES 500-503). IN
ACCORDANCE WITH THE IEC CLASSIFICATIONS (SECTION 5.5.),
THE HEARTSTART FR2+ IS NOT TO BE USED IN THE PRESENCE OF
FLAMMABLE SUBSTANCE/AIR MIXTURES.
DANGER
DO NOT RECHARGE THE M3863A FR2 STANDARD BATTERY.
WARNING
Use the HeartStart FR2+ only as described in this manual. Improper
use of the HeartStart FR2+ can cause death or injury. Do not press
the Shock button if the defibrillator pads are touching each other or
are open and exposed.
CAUTION
Hazardous electrical output. The HeartStart FR2+ is for use only by
qualified personnel.
CAUTION
Do not immerse any portion of the HeartStart FR2+ in water or other
fluids. Do not allow fluids to enter the HeartStart FR2+. Avoid spilling any
fluids on the HeartStart FR2+ or accessories. Spilling fluids into the
HeartStart FR2+ may damage it or present a fire or shock hazard. Do not
sterilize the HeartStart FR2+ or accessories.
Содержание HeartStart FR2+
Страница 4: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...
Страница 10: ...M3860A M3861A HEARTSTART FR2 P h i l i p s M e d i c a l S y s t e m s Notes ...
Страница 14: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 20: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 42: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 66: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 78: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 94: ...P h i l i p s M e d i c a l S y s t e m s M3860A M3861A HEARTSTART FR2 Notes ...
Страница 101: ...P h i l i p s M e d i c a l S y s t e m s Intentionally left blank ...