Alarm
Condition
Display
Buzzer
Remote
contact
Safety operation
Back-up of clock
function
During power failure
―――
――
―
Continuous running
by battery (CR2032)
Fan lock
Fan is locked
<Top screen>
“E07:Air circulating
fan motor trouble.”
Intermittent
tone
Alarm
status
Compressor OFF
and heater OFF
Protection of
compressor
The protective sensor is
higher than
80
o
C(MIR-154) or
85
o
C(MIR-254)
<Top screen>
“E08:Comp. is
over-heat.”
Intermittent
tone
Alarm
status
――
Time to replace
the fan
The running time is over
75,000 hours
<Top screen>
“Warning:Replace air
circulating fan mot.”
――
―
――
Disconnection of
compressor
protective sensor
The input voltage
corresponds to -50
o
C or
less disconnection of
sensor
<Top screen>
“E09:Comp.sensor is
opened.”
Intermittent
tone
Alarm
status
Compressor OFF
Defrost sensor
abnormality
The input voltage
corresponds to -50
o
C or
less disconnection of
sensor
<Top screen>
“E10:Def.sensor is
opened.”
Intermittent
tone
Alarm
status
Heater OFF
The input voltage
corresponds to 70
o
C or
more short of sensor
<Top screen>
“E11:Def.sensor is
shorted.”
Intermittent
tone
Alarm
status
Heater OFF
Time to replace
the fan motor for
compressor
The running time is over
42,000 hours
<Top screen>
“Warning:Replace
comp. cooling fan
motor”
――
―
――
Door alarm
Door is open for
2 minutes.
<Top screen>
“Door:Open”
Intermittent
tone with
delay
―
Heaters, fan motor
OFF
Power failure
alarm
At power failure.
If the power supply cord is
unplugged or the power
switch is turned off.
―――
――
Alarm
status
――