19
Caution:
The buzzer rings for a long time when a wrong password is input. Input a correct password.
The password
for key unlock must be
shared and administered by all users on this product.
The setting of key unlock when shipped from the factory is “0000”.
Refer to page 41 for changing the password.
PROCEDURES OF SETTING VALUE
Setting of Chamber Temperature
Procedures of changing chamber temperature from 25
o
C to -10
o
C are shown below. As for other values,
the same procedures are applied.
Procedures of changing chamber temperature from 25
o
C to -10
o
C
Operation Key
LCD
Display
1
----
Top screen is displayed.
2 Press the menu button (MENU).
MENU
MENU window is displayed.
3
Select Std-by in MENU window
pressing the shift key (downward).
▼
Std-by is displayed in white letters.
4 Press the enter key (ENTER).
ENTER
Stand-by Setting screen is displayed and chamber
temperature setting value is changed into white
letters.
5
Change the value from 25.0 to 10.0
pressing the character input key.
Chamber temperature setting value 10.0 is
displayed.
6 Press the menu button (MENU).
MENU
MENU window is displayed.
7
/- in MENU window pressing
the shift key (downward).
▼
+/- is displayed in white letters.
8 Press the enter key (ENTER).
ENTER
Chamber temperature setting value -10.0 is
displayed.
9 Press the enter key (ENTER).
ENTER
Next item (Light Step) is displayed in white letters.
10 Press the menu button (MENU).
MENU
MENU window is displayed. OK is already
selected (displayed in white letters).
11 Press the enter key (ENTER).
ENTER
The set value is memorized and top screen is
displayed.
CAUTION
·
The settable range of chamber temperature is between -10
o
C and +60
o
C.
The lighting is available just under the specified setting range of temperature b2
o
C and +50
o
C
when the optional light add on kit (MIR-L15) is installed. In case of setting temperature out of this range,
the lighting is not usable. The lamp will automatically turn off at temperature outside the +2
o
C and +50
o
C range when the program already saved is running.
In this case, temperature fluctuation may exceed ±1.5
o
C or it may take longer time to pull down the
chamber temperature.
0 0