TROUBLESHOOTING 

 
If the unit malfunctions, check out the following before calling for service. 
 

Malfunction Check/Remedy 

The unit does not operate   
at all. 

 The unit is not plugged correctly into a power outlet. 

 The circuit breaker at the power source is active or a power failure 

has occurred. 

The key operation is   
disable 

 The key lock function is set in ON mode. 

 

If the alarm function and 
the buzzer operates 

[At the beginning of operation] 

 The chamber temperature is not equal to the set value. 

 The chamber CO

2

 density is not equal to the set value. 

a. The secondary pressure of the gas regulator is not equal to the 
set value (0.1 MPa(G), 1.0 kgf/cm

2

(G), 14.5 psi(G)). 

b. The tube is not connected securely between the gas regulator 
and the unit. 

[During operation] 

 The upper limit alarm temperature is not set at least 1 

o

C higher 

than the set chamber temperature. 

 The set temperature value was changed, or the door was left 

open for a long period.  Or a low temperature load was placed 
inside the unit.    In this case, if the unit is left as it is, the alarm will 
eventually clear itself. 

 The gas tube has slipped off or the gas leaks. 

 The set value of the gas density was changed. 

 The gas cylinder is empty.  Check the primary pressure of the 

CO

2

 cylinder once a week.    (The primary pressure of less than 3.8 

MPa(G) (38 kgf/cm

2

(G), 551.1 psi(G)) means a little gas in the 

cylinder.   Replace the cylinder soon.) 

While an equipment are operated in chamber. (Generation of heat by 

the equipment raises temperature in chamber) 

If the chamber 
temperature is not equal to 
the set temperature 

 The ambient temperature must always be at least 5 

o

C less than 

the set temperature. 

 

If the gas density does not 
coincide with the set value 

 The secondary pressure is not set to 0.1 MPa(G) (1.0 kgf/cm

2

(G), 

14.5 psi(G)). 

• 

The gas tube is clogged or chinked. 

If the chamber humidity 
does not rise 

 The water reservoir is not filled with sterile distilled water.  

(Always use sterile distilled water.) 

 The no humidifying mode is set. 

If the CO

2

 consumption 

is too much 

 

 The door is opened frequently. 

• 

There is any gas leakage at the connection or pin hole on the 

tube.    It is recommended to replace the tube once a year.

 

• 

An access port is opened.

 

If normal cultivation 
cannot be done and 
chamber gas density is 
suspect 

• 

The environment around the unit is not normal.  The source of 

the contaminated gas is nearby.

 

• 

The unit is installed in an enclosed space.

 

If it takes much time to 
recover the gas density 

 HEPA filter is provided in the gas piping.    If it takes much time to 

recover the gas density even though the gas pressure is normal, it 
may be that dust on the HEPA filter prevents the gas flow.  
Consult the our sales representative or agent. 

If CO

gas is not 

injected 

 The CO

injection system is ON-OFF controlled and injects gas 

intermittently when the set density is close. Injection might stop for 
15 seconds but it is not abnormal.   

Содержание MCO-80IC Series

Страница 1: ...80IC Series Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 42 for model number Operating Instructions CO2 Incuba...

Страница 2: ......

Страница 3: ...CAUTIONS FOR CULTURE 18 START UP OF UNIT 19 SETTING OF CHAMBER TEMPERATURE AND CO2 DENSITY 20 KEY LOCK FUNCTION 23 ALARMS SAFETY FUNCTIONS 24 SETTING OF ALARM RESUME TIME 26 Operation after power fai...

Страница 4: ...ating instruction are subject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instruction is lost or the pag...

Страница 5: ...Symbol shows this symbol means caution this symbol means an action is prohibited this symbol means an instruction must be followed Be sure to keep this operating instruction in a place accessible to u...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...he power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire When removing the plug from the power supply outlet grip the power supply p...

Страница 8: ...e to the change of setting Be careful not to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet copy the last page when you request any repair or maintenance fo...

Страница 9: ...idity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations not to exceed 10 of the nominal voltage Other supply voltage fluctuations...

Страница 10: ...8 INCUBATOR COMPONENTS Rear upper side Upper left side Bottom of chamber When the bottom cover is removed 1 2 3 4 5 6 inner face of ceiling 7 8 9 11 10 12 13 14 15 16 17 18 19 20 22 23 21 25 24...

Страница 11: ...B are available If MCO 80GC is not used only A is available Refer to page 15 for gas cylinder connection Ensure that the gas pressure is set at 0 1 MPa G 1 0 kgf cm2 G 14 5 psi G Refer to page 39 for...

Страница 12: ...lights when the outer door is open 8 Water level alarm lamp RH PAN This lamp blinks when the water in the water reservoir is less than approximately 5 L 9 Over heat lamp OVER HEAT This lamp lights whe...

Страница 13: ...in the controller 18 Numerical value shift key Pressing this key in the setting mode causes the numerical value to shift In key lock mode pressing this key makes key lock ON or OFF 19 Digit shift key...

Страница 14: ...n between COM to N C Close between COM to N O Contact capacity DC 30 V 2 A Note When the power switch is OFF or the power failure condition the contact output between COM to N O is CLOSE the contact o...

Страница 15: ...ll the unit in stable condition to avoid the vibration or noise Unstable condition may cause vibration or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the inst...

Страница 16: ...unterclockwise to contact them to the floor Ensure the unit is level 3 Fix the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall with these hooks and rope or chain 4 Gr...

Страница 17: ...nect the gas regulator to the connecting port for CO2 gas pipe located at the rear upper side of the CO2 incubator 3 Set the CO2 pressure on the secondary side to 0 1 MPa G 1 0 kgf cm2 G 14 5 psi G at...

Страница 18: ...ut abnormal temperature rising If there is too much heat emission from the equipment in the chamber The chamber temperature may deviate very much from the setting temperature when a heat generator hea...

Страница 19: ...ially when the culture medium is spilled clean and disinfect the chamber immediately Refer to page 27 Routine maintenance for details Keep the inside panels dry To protect the inside of the unit from...

Страница 20: ...ture distribution and CO2 concentration in the chamber Stored materials Never place acid or alkaline materials or materials that release corrosive gas in the chamber Such materials can cause failure r...

Страница 21: ...the chamber and internal attachments Refer to Sterilizing of chamber and attachment on page 27 3 Place the drain tank under the unit Insert a drain outlet into the hole of drain tank 4 Fill the water...

Страница 22: ...digit shift key and numerical value shift key set the figure to 05 0 When pressed the changeable digit is shifted When pressed the figure of settable digit changes 6 Press enter key ENT Set CO2 densi...

Страница 23: ...s enter key ENT Set value is memorized The current chamber temperature is displayed Table 2 No humidifying mode setting procedure Description of operation Key operated Indication after operation 1 The...

Страница 24: ...he unit is under normal conditions The unit enters calibration mode by pressing this key Wrong key operation may affect the basic performance When the unit enters calibration mode by mistake wait for...

Страница 25: ...g change from key lock OFF to key lock ON Description of operation Key operated Indication after operation 1 The current chamber temperature and CO2 density are displayed 2 Press digit shift key for 5...

Страница 26: ...ternately with the temperature on the digital temperature indicator Gas supply line indicator blinks Intermittent tone Gas supply line is altered Remote alarm Chamber temperature sensor abnormality If...

Страница 27: ...is displayed alternately with the temperature on the digital temperature indicator Intermittent tone Remote alarm Glass frame heater abnormality If the glass frame heater or the glass frame heater rel...

Страница 28: ...tal CO2 density indicator The middle digit blinks 5 Set the figure to 020 with the numerical value shift key When pressed the figure of settable digit changes 6 Press enter key ENT The setting is memo...

Страница 29: ...ff lightly Sterilizing of chamber and attachments When the chamber of the unit is contaminated the chamber and internal attachments should be cleaned and sterilized as follows Note Take care not to da...

Страница 30: ...water reservoir with a gauze containing alcohol for sterilization and then wipe off with a dry gauze 10 Wipe the water level sensor with a gauze containing alcohol for sterilization and then wipe off...

Страница 31: ...blinking add approximately 15 L of preheated sterile distilled water at 37 o C See Fig 3 It might take approximately 2 minutes until the lamp goes off 4 Make sure the water level alarm lamp RH PAN on...

Страница 32: ...gital CO2 density indicator goes out 3 Adjust the set value to the actual value of 36 8o C with the digit shift key and numerical value shift key then press enter key ENT 4 The unit automatically reve...

Страница 33: ...en press the enter key ENT 6 The unit automatically reverts to the display mode Note This calibration is available when the setting of CO2 density is 2 or more Example For an internal CO2 density of 5...

Страница 34: ...as in the cylinder Replace the cylinder soon While an equipment are operated in chamber Generation of heat by the equipment raises temperature in chamber If the chamber temperature is not equal to the...

Страница 35: ...33 Note If the malfunction is not eliminated after checking the chart in the preceding page or the malfunction is not shown in the chart contact our sales representative or agent...

Страница 36: ...d in an unsupervised area for an extended period ensure that children do not have access and doors cannot be closed completely The disposal of the unit should be accomplished by appropriate personnel...

Страница 37: ...nformationen zur Entsorgung von elektrischen und elektronischen Ger ten private Haushalte Dieses Symbol auf Produkten und oder begleitenden Dokumenten bedeutet dass verbrauchte elektrische und elektro...

Страница 38: ...ter les autorit s locales ou votre revendeur afin de conna tre la proc dure d limination suivre Spanish Informaci n sobre la eliminaci n para los usuarios de equipos el ctricos y electr nicos usados p...

Страница 39: ...orma es sobre o m todo de elimina o correcto Italian Informazioni per gli utenti sullo smaltimento di apparecchiature elettriche ed elettroniche obsolete per i nuclei familiari privati Questo simbolo...

Страница 40: ...act op met de lokale overheid of uw leverancier en vraag wat de juiste verwijderingsmethode is Swedish Information om kassering f r anv ndare av elektrisk elektronisk utrustning privata konsumenter Om...

Страница 41: ...CO2 gas supply line A or B 6 When one cylinder is empty the indicator blinks buzzer sounds and E01 and current chamber temperature are displayed alternately on the digital temperature indicator while...

Страница 42: ...ir decreases less than approximately 5 L water is supplied little by little to prevent the water temperature from dropping rapidly The water supply valve is opened and closed repeatedly Automatic wate...

Страница 43: ...igital display CO2 controller PID control system CO2 density display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or more Water level sensor Thermal type Alarm High L...

Страница 44: ...variation 0 15 ambient temperature 25 o C setting 37 o C 5 no load Chamber humidity Normal mode Over 80 R H High humidity mode Over 90 R H Maximum heat emission 6159 6 kJ h Usable environment conditio...

Страница 45: ...tion of the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below...

Страница 46: ......

Страница 47: ......

Страница 48: ...Printed in Japan LDCL041700 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Отзывы: