INCUBATOR COMPONENTS
Remote alarm terminal
The remote alarm terminal is located at the upper
left side of the unit.
The remote alarm terminal is a contact output.
Normal : Open between COM to N.O.
Close between COM to N.C.
Abnormal: Open between COM to N.C.
Close between COM to N.O.
Contact capacity : DC 30 V, 2 A
Note:
• When the power switch is OFF or the power failure condition, the contact output between COM to N.O.
is CLOSE (the contact output between COM to N.C. is OPEN).
• The remote alarm cannot be silenced by pressing the alarm buzzer stop key (BUZZER) since the
remote alarm is not conjunct with the alarm buzzer stop key (BUZZER).
Remote Alarm
MAX. DC 30 V, 2 A
N.C.
COM
N.O.