16
CAUTIONS FOR USING EQUIPMENT IN CHAMBER
●
Avoid abnormal temperature increase when an equipment is used in the chamber
When using a heat emitting equipment such as a shaker or a bottle roller, the chamber temperature may
rise abnormally. Abnormal rising of the chamber temperature is dangerous, so make sure to operate
equipment without abnormal temperature rising.
●
If there is too much heat emission from the equipment in the chamber
The chamber temperature may deviate very much from the setting temperature when a heat generator
(heater, motor, etc.) of the equipment is energized in the chamber.
●
When operating an equipment in the chamber, use at "No humidifying mode" (refer to page 21).
High humidity may cause a failure, a current leakage or a fire to an equipment in the chamber.
●
When operating an equipment in the chamber, make sure there is no water in the water
reservoir.