52
H
2
O
2
DECONTAMINATION (OPTIONAL)
11.
Press “H
2
O
2
” for 3 seconds to open the Setting
Position window. Enter the password if prompted.
12.
Press
“Next”
to
display
the
H
2
O
2
Decontamination Step1 screen. If there is
something to worry about setting position, check
the inside again.
13.
The system check starts automatically. If an
alarm or safety function is activated, refer to Table 8
on page 66.
14.
If the system is normal, the H
2
O
2
Decontamination Step2 screen appears. Press
“OK” to start H
2
O
2
decontamination. H
2
O
2
decontamination is performed automatically from
Step3 to Step8 (procedure
15
).
Notes:
• The outer door is electrically locked for safety until
H
2
O
2
decontamination is completed.
• A buzzer sounds when H
2
O
2
decontamination is
completed (after about 100 minutes).