51
7.
Remove the connector cap from the connector
on the bottom right of the far side of the chamber.
Connect the H
2
O
2
generator and connector by the
cable provided (Fig. 2).
Note:
Be sure to keep the connector cap.
8.
Insert four trays in the 2nd, 3rd, 4th, and 6th tray catches from the top of the chamber side.
Note:
The trays included as accessories are designed to be appropriate for decontamination. If half trays
(MCO-35ST, optional) or trays for previous models are used, decontamination may not be sufficiently
effective.
9.
Insert the humidifying pan cover and humidifying pan removed in procedure
1
(Fig. 3).
10.
Make sure that the duct, the fan, and the H
2
O
2
generator are securely attached (refer to page 24).
Then close the inner and outer doors.
Fig. 3
Front
Humidifying
pan cover
Humidifying pan
H
2
O
2
generator
Fig. 2
Connector
Front