P118i Ultrasonic Flowmeter and Analyzer
Reversion: 3.0.0
Page 31 of 49
Fluid Sound Velocity
Enter the fluid sound velocity. It only can be used
when item "Other" is selected in window M20, i.e. it
is unnecessary to enter all the fluids listed in Window
M20.
Fluid Viscosity
Enter fluid
’
s kinematics viscosity. It only can be used
when item "Other" is selected in Window M20, i.e. it
is unnecessary to enter all the fluids that listed in
Window M20.
Transducer Mounting
Three mounting methods are available:
0.
V
1.
Z
2.
N (small pipe)
Transducer Spacing
The operator must mount the transducer according to
the transducer spacing displayed (be sure that the
transducer spacing is measured precisely during
installation). The system will display the data
automatically after the pipe parameter has been
entered.