P118i Ultrasonic Flowmeter and Analyzer
Reversion: 3.0.0
Page 37 of 49
Scale Factor
The scale factor is used to modify the measurement
results. The user can enter a numerical value other
than "1" according to calibration results.
Network IDN
Input system identifying code, these numbers can
be selected from 1~247 except that 13 (0DH
ENTER), 10 (0AH Newline), 42 (2AH
﹡
) and 38
(26H&) are reserved. System IDN is used to identify
the flowmeter to a network.
System Lock
Lock the instrument.
Once the system is locked, any modification to the
system is prohibited, but the parameter is readable.
"Unlock" using your designated password. The
password is composed of 6 numbers.
Segmented Correction
ON: Turn on the Sectional Correction Function;OFF:
Turn off the Sectional Correction Function
(
optional
)
Only when it is "ON", the settings of the Sectional
Correction Value in M49 will be workable.