60
BG 5195 BEN / A (2011-05) TPG261.oi
Test of the main memory.
Test sequence
The test runs automatically one
time:
Ö
Test in process (very briefly).
Ö
Test finished, no error found.
Ö
Test finished, error(s) found.
The
lamp flashes.
If the error message persists
after several test sequences
have been run, please contact
your local Pfeiffer Vacuum
service center.
Test of the program memory.
Test sequence
The test runs automatically one
time:
Ö
Test in process
Ö
Test finished, no error found.
After the test, a four-digit
checksum (hexadecimal for-
mat) is displayed.
Ö
Test finished, error(s) found.
After the test, a four-digit
checksum (hexadecimal for-
mat) is displayed. The
lamp flashes.
If the error message persists
after several test sequences
have been run, please contact
your local Pfeiffer Vacuum
service center.
RAM test
EPROM test