BG 5195 BEN / A (2011-05) TPG261.oi
83
Transmit:
UNI
[,x] <CR>[<LF>]
│
└─
Pressure unit, x =
0 –> mbar/bar (default)
1 –> Torr
2 –> Pascal
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Pressure unit
Transmit:
BAU
[,x] <CR>[<LF>]
│
└─
Transmission rate, x =
0 –> 9600 baud (default)
1 –> 19200 baud
2 –> 38400 baud
As soon as the new baud rate has been en-
tered, the report signal is transmitted at the new
transmission rate.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Transmission rate
Transmit:
DCD
[,x] <CR>[<LF>]
│
└─
Resolution, x =
2 –> Display x.x (2 digits)
(default)
3 –> Display x.xx (3 digits)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
│
└─
Resolution
When the PrE (
→
46) is ON and the pressure is in the
range p<1.0E-4 mbar the display resolution of the PCR
Gauge is reduced by one decimal digit.
5.2.2.4 General
Parameters
Pressure unit
Transmission rate
Display resolution