BG 5195 BEN / A (2011-05) TPG261.oi
11
Number 2
Reaction delay
≤
20 ms if switching threshold
close to measurement value
(for larger differences con-
sider filter time constant)
Adjustment range
depending on gauge
(
→
[1] … [17])
Hysteresis
≥
1% F.S. for linear gauges,
≥
10% of measurement value
for logarithmic gauges
Contact type
floating changeover contact
Load max.
30 VAC, 30 W (ohmic)
60 VDC, 1 A, 30 W (ohmic)
Service life
mechanic
electric
5×10
7
cycles
1×10
5
cycles (at max. load)
Contact positions
→
22
Relay
connector
D-Sub appliance connector,
female, 15-pin
(pin assignment
→
22)
Number 1
Reaction time
≤
20 ms
Contact type
floating normally open contact
Load max.
30 VAC, 30 W (ohmic)
60 VDC, 1 A, 30 W (ohmic)
Service life
mechanic
electric
5×10
7
cycles
1×10
5
cycles (at max. load)
Contact positions
→
21
Control
connector
Amphenol C91B appliance
connector, female, 7-pin
(pin assignment
→
21)
Switching functions
Switching function relays
Error signal
Error signal relay