21
Feature
Unit
TPH 2201
TPH 2201
Connection nominal diameter
Inlet
DN 250 ISO-K
DN 250 CF-F
Outlet
DN 63 ISO-K
Venting connection
G 1/8”
Nominal rotation speed
1/min
36 000
Standby rotation speed
1/min
24 000
Start-up time (up to 90% of the nominal
rotation speed, fore-vacuum pressure
≤
0,1 mbar)
min
13
Noise level
dB (A)
59
Maximum permissible rotor temperature
°C
120
Permissible heat radiation power
W
30
Volume flow rate for:
Nitrogen N2
l/s
2100
Helium He
l/s
2500
Hydrogen H2
l/s
2000
Compression ratio for:
N2
108
He
4 · 104
H2
2 · 103
Maximum fore-vacuum pressure
N2
mbar
0,25
Argon (Ar)
mbar
0,23
Maximum gas throughput 1) with water cooling
N2
mbar l/s
5,4
Ar
mbar l/s
2,7
Vertex power characteristics lines2)
A
W / Hz
560 / 600
B
W / Hz
560 / 600
C
W / Hz
560 / 600
D
W / Hz
560 / 600
Final pressure3)
with rotary vane vacuum pump
mbar
< 1 · 10-9
Lubricant
Type4)
TL 011 oder F3
Filling volume
cm3
2 x 125
Cooling water consumption
with water at 15 ˚C5)
l/h
40
Cooling water temperature
˚C
5 - 25
Permissible magnetic field
mT
13,4
Power consumption heating unit
W
400
Weight
kg
79
84
Connection voltage
VDC
140 ± 5%
Current consumption
A
4,3
Power
W
600
Protection class
IP 30
9. Technical Data
1) Measured with a rotary vane vacuum pump > 60 m3/h,
higher gas throughputs with reduced rotation speed.
2) For gas type characteristics lines please refer to
Section 4.7.
3) In accordance with DIN 28 428 the final pressure of a
turbomolecular pump is that pressure which is attained in
a measuring dome 48 hours after baking out.
4) See rating plate.
5) With maximum gas throughput up to 100 l/h.