6. What To Do In Case Of Breakdowns?
Problem Possible
Causes
Remedy
Pump doesn't start;
• Power supply interrupted
• Check fuse in the power pack unit
None of the integrated LEDs
• Check plug contacts on the power pack unit
glow on the TC 600
• Check power pack unit feeder line
• Check voltage on the power pack
unit (140V DC) at connection X2
• Incorrect operations voltage supplied • Supply correct operations voltage
• Pins 1-3 and 1- 4 on the remote-
• Connect pins 1-3 and 1- 4 on the remote plug
plug not connected
• No supply of operations voltage
• Check plug contacts on the power pack unit
• Defect TC 600
• Inform PFEIFFER Service of need for repair
• Reduction in the voltage in the cable
• Use suitable cable
Pump doesn't attain nominal
• Fore-vacuum pressure too high
• Check backing pump function
rotation speed within the set -
• Leak or too much gas
• Check seals
start-up time;
• Seek leak and repair
Pump cuts out during
• Supply of process gas too high; reduce
operations
• Rotor sluggish caused by
• Check bearing (noises?): Request
defective bearing
PFEIFFER Service to replace
• TC start-up time too short
• Set longer start-up time with the DCU or PC
• Thermal overloading caused by:
– Insufficient water flow
• Ensure free flow
– Insufficient air supply
• Ensure adequate air supply
– Fore-vacuum pressure too high
• Reduce fore-vacuum pressure
– Ambient temperature too high
• Reduce ambient temperature
• Leak or too high a gas load
• Seek leak in the system and repair
• Process gas supply too high; reduce
• Lubricant or lubricant
pump dirty
• Carry out lubricant change as per
Section 8.1. or clean the lubricant
pump as per Section 8.3.
• Lubricant deficiency
• Check lubricant and replace as necessary
Pump doesn't attain final pressure
• Pump dirty
• Bake out pump
• If seriously contaminated: Request
PFEIFFER Service to clean
• Leak in vacuum chamber,
• Seek leak starting with vacuum chamber
lines or pump
• Repair leak
Unusual operating noises
• Bearing damaged
• Inform PFEIFFER Service of need for repair
• Rotor damaged
• Inform PFEIFFER Service of need for repair
• Splinter shield (if fitted) not
• Check seat of splinter shield
seated firmly
(see Section 3.2.)
Red LED on the TC 600 glows
• Collective malfunction
• Reset via mains OFF/ON or remote pin 5
• Different malfunction display
with a DCU possible
Red LED on the TC 600 blinks
• Warning from:
• Different warning display
with a DCU is possible
– Mains power failure
• Ckeck power pack voltage
• Check the power pack mains connection
– Supply voltage short
circuit to earth
• Check the power pack unit voltage for short
circuit to earth