10
3.11. Connecting The Serial
Interface RS 485
An external operating component (DCU 001/DCU 600) or an
external computer can be connected via the connection
”RS 485” on the TC 600 with the use of an 8 pole modular
connecting cable.
The serial interface is galvanically and safely separated from
the maximum supply voltage from the TC 600.
Connection
Connecting The RS 485
Connecting to a fixed bus system:
➡
Connect all units with D+ (pin 5/RS 485) and
D- (pin7/RS 485) to the bus.
➡
The bus must be connected at both ends.
The connections should be made in accordance with the
specification of the Serial Interface RS 485.
PIN
Arrangement
1
not connected
2
+24V output (
≤
210 mA loading
capacity)
3
not connected
4
not connected
5
RS 485: D+ (DO / RI)
6
Gnd
7
RS 485: D- (DO / RI)
8
not connected
All units connected to the bus must have differing serial inter-
face addresses (parameter 797).
The group address of the TC 600 is 960.
Only safety low voltages (SELV) may be connec-
ted to Serial Interface RS 485 .
All switched on remote functions have priority over the serial
interface functions.
Please Note:
Please refer to the separate Operating Instructions
PM 800 488 BN for more detailed information on operations
via Serial Interface RS 485 and its electronic data.
1 ... 8
Description
Value
Serial Interface Type:
RS 485
Baudrate:
9600 baud
data word lenghts:
8 bit
Parity:
no (no parity)
Startbits:
1
Stopbits:
1..2
The electrical connections are internally optically decoupled.
Please note:
It is possible to connect an RS 232 (e.g. PC) via a level conver-
ter (please see ”Accessories”).
(View from the
plug side of the
TC 600)
RS 485
5:D+, 7:D–
R
T
R
T
PC
Level Converter
TC
TC