3 Installation
Note
Store the transport fastening devices and re-install them before
transporting the product again.
DANGER
Voltages up to 330 VDC appear on the QMA 200 during operation.
Other parts in the vacuum chamber (e.g. gauges) may possibly be exposed to
this voltage under unfavorable conditions. If as a result such parts (including the
lines and connected equipment!) become hazardous, they must be arranged or
protected in such a way that no sparkover and charge transfer can occur.
If the QMA can be touched while the vacuum system is open, additional
protection is needed, e.g.:
•
mechanical protection against touching
•
forced disconnection of the SP 200 power supply when the system is opened.
DANGER
No external hazardous voltages are allowed to be transferred to the electrodes of
the QMA 20. If such danger sources exist in the vacuum chamber, additional
safety measures (e.g. better placement, shielding, grounding, etc.) must be taken
that reliable preclude such influences. Also lower external voltages acting on the
QMA may damage the electronics and lead to unreliable measurement results.
Suitable precautions are to be taken as mentioned above.
WARNING
The QMA must be correctly installed from a vacuum engineering point of view,
i.e. the gases to be detected must have unhindered access to the analyzer.
Vacuum coating of the analyzer must be prevented.
Note
Make sure you adhere to the special requirements concerning cleanliness and
damage prevention when working with vacuum components.
3.1 Unpacking
3.2 Montage