![Pfeiffer Vacuum Prisma QMS 200 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/pfeiffer-vacuum/prisma-qms-200/prisma-qms-200_operating-manual_1534985011.webp)
BG 805 201 BE (9911) QMS200.om
11
LAN
Pin
Assignment
JIS F07
(TOSLINK)
T
Transmit
R
Receive
RS-232-C
Pin
Assignment
Pin
Assignment
D-Sub.
9-pin male
1
n.c
6
n.c
2
Received Data (RXD)
7
optional signal
Request To Send (RTS)
3
Transmitted Data (TXD)
8
optional signal
Clear To Send (CTS)
4
optional signal
Data Terminal Ready
(DTR)
9
n.c
5
Signal Ground (SGND)
Case
Shield
WARNING
When this connection is used a shielded cable is required for EMC reasons. The
shield must contact the connector housing. The opposite end should remain
open or be grounded so that no damaging compensating currents can occur.
24 V DC
Pin
Assignment
3-pin female
1
+ 24 V DC
2
GND
3
GND
Case
Screen
QMA
Pin
Assignment
13-pin, male
1
Reserve
2
Extr
3
Focus
4
HV -
5
Filament 1
6
Filament 2
7
Anode
8
RF -
9
GND
10
GND
11
RF +
12
Filament Common
13
EP
DANGER
This connection carries hazardous
voltages
tune
control
LAN
ready
24V
RS-232-C
VACUUM
tune
control
LAN
ready
24V
RS-232-C
VACUUM
T
R
6
1
9
5
1
2
3
6
7
8
9
10
11
13
12
1
2
3
4