3.8. Connecting The Pump To The

Electronic Drive Unit TCP 2000

Electrical connections must be carried out in
accordance with local regulations. Voltage and
frequency values on the TCP 2000 rating plate
must concur with mains voltage and mains
frequency values.

Connect the terminal box 53 with the flanged-on X1 round
plug into the mating plug on the pump.

Connect the signal wire to the D-Sub connector at the X3B
socket of the TCP 2000.

Connect the motor cable 52 to the power terminal X1A

WARNING

When the operating voltage is provided and the

main switch switched "ON" the pump runs a self
test so as to check the supply voltage.

Only unplug the connector to the electronic
drive unit when the pump is completely at rest
and the electronic drive unit has been 
disconnected from the mains.
Voltages of more than 130 Veff can occur on a

running down pump on the open electric
connection.
There is a danger of an electric shock from
touching the contacts.

For further details please refer to Operating Instructions 
PU 0056 BN for the Electronic Drive Unit TCP 2000.

WARNING

CAUTION

U

V

W

95080402

07.F5.S0C-1220

L1

L2

L3

+

-

PB

U

V

W

X1

X2

X1A

U

V

W

X3B

X3

Connecting The Pump To The Electronic
Drive Unit TCP 2000

P

OnTool

TM

Booster

50 Electronic Drive Unit TCP 2000
52 Motor connecting cable
51 Mains connection
53 Terminal 

box

X1A Power terminal
X2 Connecting cable cooling water 

monitor

X3 Connecting cable for inert gas valve
X3B Connector for signal wire

Controlling the direction of rotation

Exchanging two phases when attaching the pump

cable leads to a wrong direction of rotation.
Pay attention to the markings (U-U/V-V/W-W) at
the pump cable when attaching the pump.

CAUTION

50

53

52

Characteristics for the correct direction of rotation:

Open vacuum flange.

Switch on pump;
– in case of correct direction of rotation an air flow esca-

pes at the exhaust flange which is decreased after
sealing of the vacuum flange.

– the blank flange cannot be loosened after sucking in at

the vacuum flange.

Содержание OnTool Booster 150

Страница 1: ...Translation of the original instructions Operating Instructions EN PU 0058 BEN 1 OnTool BoosterTM 150 Side Channel Pump with TCP 2000...

Страница 2: ...c Drive Unit TCP 2000 11 3 9 Using Sealing Gas 12 3 10 Pumping Of Vapors 13 3 11 Pumping Of Light Or Heavy Gases 13 4 Operations 14 4 1 Before Switching On The Pump 14 4 2 Switching The Pump ON 14 4 2...

Страница 3: ...between the pump and the TCP 2000 during operations When the vacuum flange is open disconnect the voltage supply to the TCP 2000 When working on the pump the vacuum flange should only be opened once t...

Страница 4: ...TE Danger of burns from touching hot parts WARNING Danger of an electric shock WARNING Danger of injury from rotating parts WARNING Danger of personal injury WARNING 1 2 Pictogram Definitions Danger o...

Страница 5: ...ax permissible intake pressure for continous operation 1000 mbar for N2 and air Max permissible vacuum chamber volume for loadlock applications 1 m 3 Improper Use The following are regarded as imprope...

Страница 6: ...zzles for the cooling water monitor and the cooling water connection input output including sealing rings Reducing nipple for the cooling water monitor Cooling water monitor Blank flange DN 50 ISO KF...

Страница 7: ...suitable transport straps or screw an eye bolt into the transport bores T and lift the pump with lifting bar Ensure even distribution of weight when lifting Danger of tilting WARNING WARNING 3 3 Sett...

Страница 8: ...ot fill when the pump is running and not in evacuated condition Unscrew both lubricant filler screws 31 taking care with the sealing Slowly fill in 24 ml of the accompanying lubricant into each lubric...

Страница 9: ...nclean components prolong the pumping time Fit a vacuum safety valve Accessory in the vacuum line This prevents the vacuum chamber being ventilated via the backing pump PLEASENOTE Cover and bolt down...

Страница 10: ...deposits collecting in the pump the cooling water must be filtered Minimum cooling water requirements Mechanically clean optically clear no turbidity no sediment chemically neutral Oxygen content max...

Страница 11: ...s a danger of an electric shock from touching the contacts For further details please refer to Operating Instructions PU 0056 BN for the Electronic Drive Unit TCP 2000 WARNING CAUTION U V W 95080402 0...

Страница 12: ...ugh between the suction chamber and the bearing chamber on the exhaust side This effects the following protection of the lubricant against contamination protection of the bearing when pumping vapors a...

Страница 13: ...the case of lighter gases the gas ballast operates as a carrier gas which promotes pumping of the process medium If gases other than air or nitrogen are used for the gas ballast please discuss such pl...

Страница 14: ...change of lubricant the contacts for monitoring the lubricating pump can open as a result of degassification In such cases as for other malfunctions the electronic drive unit switches off the pump PLE...

Страница 15: ...rocess has been completed allow the pump to continue running for about 30 minu tes with open gas ballast valve for operating fluid regeneration purposes CAUTION 4 4 Switching The Pump OFF Details rega...

Страница 16: ...ve unit defect Check ventilator function Cooling water monitor defective or not connected Check cooling water monitor The pump does not attain the final Dirty pump In serious cases inform Pfeiffer Vac...

Страница 17: ...ess related toxic gases and vapours can escape from the lubricant which may become comtani mated with harmful substances radioactive chemical etc Please request safety instruction data sheets for lubr...

Страница 18: ...position Do not fill when the pump is running and not in evacuated condition Unscrew both lubricant filler screws 31 taking care with the sealing Slowly fill in 24 ml of the accompanying lubricant int...

Страница 19: ...vely or radioactively cannot be accepted as a matter of principle Fill out the service request and the declaration on contamination Download the forms Service Request and Declaration on Contamination...

Страница 20: ...ant F3 Amount of lubricant for the lubricant reservoirs ml each 24 Cooling water consumption with water at 15 C l min 2 Cooling water temperature C 15 30 Permissible ambient temperature C 12 40 Weight...

Страница 21: ...Elbow union with orifice G 1 8 PR 003 578 U orifice 0 2 mm Electronic drive and control unit TCP 2000 PR C01 000 Control panel 1 phase Electronic drive and control unit TCP 2000 PR C01 001 Control pan...

Страница 22: ...22...

Страница 23: ...nsible for compiling the technical documentation is Mr Sebastian Oberbeck Pfeiffer Vacuum GmbH Berliner Stra e 43 35614 A lar OnTool BoosterTM 150 Guidelines harmonised standards and national standard...

Страница 24: ...e the only supplier of vacuum technology that provides a complete product portfolio Benefit from our know how and our portfolio of training opportunities We can support you with your plant layout and...

Отзывы: